Calibration
WTX120
A4500-1.1
HBM: public
117
11
Calibration
To access the
Calibration
group select the following keys:
Menu
Key
Function
Standard
Switch from
Standard
to version notice
WTX120V#.#
Display the current firmware version number
Call password entry within 3 seconds
Password
324
Enter the password
1 ServiceMode
Call
ServiceMode
Service:
Interface
Press twice to switch to the
Calibration
group
options
Service:
Calibration
11.1
Multiple-Range
The A/D converter can be configured as a single, dual or triple-range scale.
The individual ranges differ in their maximum load and scale interval.
Example of three-range scale with a resolution of 3000 increments in each
weighing range:
Lowest weighing range: 0 to 1500kg / 0.5 kg
Middle weighing range: 0 to 3000kg / 1.0 kg
Highest weighing range: 0 to 6000kg / 2.0 kg
The multiple-range scale automatically switches from the lower range to the
next higher range when under load.
Summary of Contents for WTX120
Page 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Page 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Page 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Page 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...