WE2107
20
A2087-1.3 en/de
HBM
LEGAL parameter:
LEGAL setting
Display, lower limit value
Display, upper limit value
Not legal‐for‐trade
-160 %
160 %
OIML
-20d
CAP + 9 d
NTEP
-2 %
105 %
LEGAL setting
Tare range, lower limit
Tare range, upper limit
Not legal‐for‐trade
-100 %
100 %
OIML, NTEP
>0
100 %
The % figures relate to the nominal weighing range (“AdJ
³
CAP” parameter)
Each time this parameter is changed, the legal‐for‐trade counter is incremen
ted. This is displayed in the Information sub‐menu.
5.4.3 Setting and adjusting the scale
This function can only be executed when the hidden pushbutton (for calling
the parameter menu) is pressed.
Access
level
Main
menu
level
1
Second
menu
level
2
Third
menu
level
1 2
Type
Sub‐menu explanation
1 2
(
flashing) = parameter input /
selection
4
“AdJ
³
Scale adjustment parameters
SEtUP
³
Unlt”
S
Unit selection:
OFF / g / kg / t / lbs
AZEro”
S
Zero on startup:
OFF /
±
2 / 5 / 10 / 20
ZtrAc”
S
Automatic zero tracking:
OFF / ON (0.5 d/s)
StiLL”
S
Standstill monitoring:
±
OFF / 0.5 / 1 / 2 / 5 d/s
rES”
S
Increment:
1 / 2 / 5 / 10 / 20 / 50 d
Point”
S
Decimal point:
0 = xxxxx.
1 = xxxx.x
2 = xxx.xx
3 = xx.xxx
4 = x.xxxx
CAP”
I
Max. capacity (weighing range):
100
…
99999
rAnG1”
I
Changeover point for dual‐range
balance:
0
…
99999 of the nominal (rated)
value,
0 = Single‐range balance