5.3
Installation for tensile loading/compressive loading
5.3.1 Installation without adapter
The transducer is screwed directly on to an existing structural element (e.g.
profile, cover, plate). This type of installation enables the transducers to
measure axial forces in the tensile force and compressive force directions.
Alternating loads can also be recorded perfectly. To achieve this, the
transducer must be mounted without any axial play, which is made easier by
centering aids on the top and bottom of the transducers.
Usable centering assistance on the top is 2.5mm and on the bottom, 1.5mm
(see page 24).
+
The structural elements provided by the customer must be parallel to one
another. The screws must be tightened in a diagonally opposite sequence.
Section drawn rotated at the plane of projec-
tion
Installation with flange through-hole
Installation with flange thread
provided by
customer
provided by
customer
M
2
(M
A
)
M
1
(M
A
)
U5
M
3
(M
A
)
M
2
(M
A
)
Fig. 5.1
Installation for tensile loading/compressive loading
Nominal
force (kN)
Starting tor-
que M
A
(Nm)
Screws for transducer mounting M
1
(flange
through-hole)
Resistance class 10.9
metric
UNF
T
(mm)
100
115
8 x M12
8x1/2”
approx. 15
200
280
8 x M16
8x5/8”
approx. 19
500
560
8 x M20
8x3/4”
approx. 23
Nominal
force (kN)
Screws for transducer mounting M
3
(flange
thread) Resistance class 10.9
Screws M
2
Resistance class 10.9
100
8xM12
8 x M12
200
8xM16
8 x M16
500
8xM20
8 x M20