Mechanical installation
22
A03452_15_X00_00
HBM: public
T40B
Important
Use a threadlocker (medium strength, e.g. LOCTITE) to glue the screws into
the counter thread to exclude prestressing loss due to screw slackening, in the
event of alternating loads.
Notice
Comply with the minimum thread reach as per table 5.1. The maximum thread
reach must be selected so that the counter flange is not touched. Otherwise
significant measurement errors may result from a torque shunt, or the trans
ducer may be damaged.
Measuring
range
Fastening screws
Prescribed
tightening
moment
Minimum
thread reach
N
V
m
Property class
N
V
m
mm
50
M8
10.9
34
100
M8
10.9
34
200
M8
10.9
34
500
M10
10.9
67
1 k
M10
10.9
67
2 k
M12
10.9
115
3 k
M12
12.9
135
5 k
M14
12.9
220
10 k
M16
12.9
340
1)
DIN EN ISO 4762; black/oiled/
m
tot
= 0.125
2)
d = screw diameter in mm
Tab. 5.1