11
MGCplus−ML71BS6
A0851-3.1 en
HBM
4.1.1
Panel output signals CAN 1
This is where to set the sources for the signals to be transmitted and if
necessary, scaling for the output values.
1
Use the shift key
to change to set−up mode and then press the
function key
. Use the cursor keys to select “Sources CAN1” and
confirm with
.
Format 1; Absolute values Float
Sources CAN1
CHANNEL 3
CAN signal 1
System
Display
Parameters
Options
Gross
Source for
MGC channel
Signal
CAN signal 1
CAN signal 2
.
.
CAN signal n
Gross
Net
Store1
Store2
Combined PV
8
Fig. 4.2:
CAN1 / CAN 2 output signals dialogue window
2
Use the cursor keys
to select the “Source for MGC channel” selection
field. Use
to call the selection list and choose the desired CAN signal.
Confirm with
.
3
Use the cursor keys
to select the MGC channel field and enter the
desired channel (e.g. 8.1).
4
Use the cursor keys
to select the “Signal” edit field. Use
to call the
selection list and choose the desired signal. Confirm with
.