B.11
Measurement cables
B.11.1
Shielded 2 Wire 600 V RMS CAT II cable
KAB290: Shielded 2 Wire 600 V RMS CAT II Cable (Option, to be ordered separately)
This cable is specially designed to be used with the GN610, GN611 and GN610B, GN611B cards. Significantly reduces signal disturbance
pickup by using two identical signal wires with earthed shield. This cable must not be used for 3 wire measurements. The shield is not a standard
signal wire.
Cable setup
2 wires with shield and isolation
3 shrouded banana plugs on each side: red, black and yellow
2 wires 0.75mm
2
Red and black
Isolation
Wire to wire
Shield to wire
Shield
Shield yellow
Cable isolation
Figure B.77:
Shielded cable setup
Wire thickness
2 * 0.75 mm
2
(0.00116 in
2
)
Maximum wire resistance
0.250 Ω / m (0.0763 Ω / ft)
Weight
Approximately 143 g/m (1.54 oz/foot)
Outside cable diameter
Approximately 9 mm (0.354 inch)
Minimum bend radius
10 times that of the cable diameter
Isolation
Resistance
20 MΩ / km (32.19 MΩ/ mile)
Voltage
600 V RMS CAT II; wire to wire; wire to shield; shield to outside
Capacitance
Wire to wire
Approximately 110 pF/m (33.54 pF/ft)
Wire to shield
Approximately 150 pF/m (45.73 pF/ft)
Temperature range
Operational
-
15 °C (+5 °F) to +80 °C (+176 °F)
Non
-
operational (storage)
-
40 °C (
-
40 °F) to +80 °C (+176 °F)
Available lengths
1.5 m (4.92 ft), 3.0 m (9.84 ft), 6.0 m (19.7 ft)
GEN3i
786
I3763-3.1 en HBM: public