Functions and outputs
Quantum
X
I3031-14.0
HBM: public
179
RMS
+
1
T
*
ŕ
T
0
f
(
x
)
2
dx
Ǹ
Where f(x) denotes the input channel signal and T the
time window (in ms).
Keep in mind the following notes when working with RMS
channels:
S
The maximum sampling rate is 4800 (2400) Hz
S
The output (sampling) rate of the RMS channel must
not be higher than the sampling rate of the input chan
nel.
S
Filters set for the MX410B do not apply to RMS chan
nels. These channels are always unfiltered. However,
the input signal is filtered.
10.2 MX460B
The MX460B supports 4 special math channels for analy
sis of rotational parts: Rotational vibration and angle dif
ference.
Fig. 10.2
MX460B front view