47
DSE
WEB INTERFACE
Net display
Shows the current net load on the scale in orange. The display also
warns you if the scale is overloaded.
The measured net value is available in object “0x601” (I16) or
“0x301a” (F32).
Gross display
The scale's gross value is displayed in blue.
The measured gross value is available in object “0x6144” (I16) or
“0x3144” (F32).
Graph
The graph depicts the progression of the load on the scale over the
last 10 seconds, differentiating between gross (blue) and net
(orange). In the Filler and Checkweigher application modes, the graph
shows the result of the last operation.
Zero
Use ZERO to balance the scale to zero.
Please note:
Zeroing is only possible when the scale is at a standstill. To cancel
the action, click the button again.
The scale can only be zeroed within a range of ±20% of the maximum
capacity. The positive value of this range is available in object
“0x6140_01” (I16) or “0x3140_01” (F32). The negative value of this
range is available in object “0x6140” (I16) or “0x3140” (F32).
Zero with object “0x6002” and command “0x6f72657a”. The zero
value is available in object “0x6142” (I16) or “0x3142” (F32).
Tare
Clicking on TARE calculates the tare value. The current tare value is
displayed above the button.
The gross value is the sum of net and tare.
Tare with object “0x6002” and command “0x65726174”. The tare
value is available in object “0x6143” (I16) or “0x3143” (F32).
Measurement
sequence
Here the gross and net values of the last approximately 10 seconds
are displayed, with their time stamp. In the Filler and Checkweigher
application modes, the results of the most recent operations are
displayed.
The content from the Home view is shown with many menus, so you can always see the
current measured value.
7.2
Device
This menu displays information about the device such as its serial number and firmware
version. You might need this information if you have problems operating the DSE and are
in contact with HBM Technical Support (see section 11, page 97). You can also see a
number of values from the device: various temperatures, the humidity in the housing, and
the current supply voltage. Changes are not possible here.
Summary of Contents for DSE
Page 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Page 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Page 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Page 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Page 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Page 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Page 203: ...DSE ENTSORGUNG 102 ...
Page 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...