Important
In the event of voltage surges on the DC system (as per 61000-4-5 and EN 45501) the
DSE module may restart, causing it to fail sporadically.
The module's downtime is usually about 1 second.
The time taken to reconnect to a PROFINET master depends on the connected network.
A time of 10 to 20 seconds is typical.
If you are setting up a weighing system or scale in accordance with OIML R 76:2006, you
must take account of this malfunction and indicate it by a relevant display.
5.3
Status LEDs
PROFINET LEDs (left: BF, ERR, NS, BE and right: SF, RUN, MS, BS)
See also Fig. 5.1 on page 14 concerning the location of the LEDs.
BF LED
Status
Meaning (bus error LED)
Off
No error.
Flashing (2 Hz)
No data exchange.
On
Error: No configuration, slow physical connection or no
connection.
SF LED
Status
Meaning (system error LED)
Off
No error.
Flashing
(1 Hz, 3 sec.)
A DCP signal service is triggered via the bus.
On
Watchdog timeout:
There is a system error, or a channel-specific, generic or
extended diagnostic.
System LED (status)
DSE LED
Device status
Meaning (system error LED)
On
The system is working without error, within the
specification.
Flashing (1 Hz)
Flashes to identify the DSE.
Summary of Contents for DSE
Page 1: ...DSE ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Page 3: ...DSE ENGLISH DEUTSCH Operating Manual ...
Page 22: ...DSE INSTALLATION 20 Fig 5 7 DSE in star topology left and ring topology right ...
Page 41: ...39 DSE CONNECTINGDSE VIA TCP IP Fig 6 24 DSE signals ...
Page 102: ...DSE ENGLISH DEUTSCH Bedienungsanleitung ...
Page 122: ...21 DSE INSTALLATION Abb 5 7 DSE in Sterntopologie links und Ringtopologie rechts ...
Page 203: ...DSE ENTSORGUNG 102 ...
Page 204: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05529 01 X00 02 ...