11
C5
CONDITIONS ON SITE
6
CONDITIONS ON SITE
Protect the transducer from weather conditions such as rain, snow, ice, and salt water.
6.1
Ambient temperature
The effects of temperature on the zero signal and rated output are compensated. To
obtain optimum measurement results, comply with the nominal (rated) temperature
range. The strain gages are designed and arranged to ensure a high insensitivity to
temperature gradients. Despite this, temperatures that are constant and, ideally, change
slowly have a favorable effect on accuracy. A radiation shield and thermal insulation on
all sides provide noticeable improvements
,
but must not form a force shunt.
6.2
Moisture and corrosion protection
Regardless of the choice of electrical connection, the sensors achieve IP 54 protection
according to DIN EN 60529 (splash-proof).
We recommend protecting the sensor from long-term exposure to moisture and weather
conditions.
6.3
Deposits
Dust, dirt and other foreign matter must not be allowed to accumulate to such an extent
that some of the measuring force is diverted, invalidating the measured value (force
shunt).
Summary of Contents for 10M0
Page 1: ...C5 ENGLISH DEUTSCH FRAN AIS Mounting Instructions Montageanleitung Notice de montage...
Page 3: ...C5 ENGLISH DEUTSCH FRAN AIS Mounting Instructions...
Page 28: ...C5 VERSIONS AND ORDERING NUMBERS 26...
Page 29: ...C5 ENGLISH DEUTSCH FRAN AIS Montageanleitung...
Page 49: ...21 C5 EINBAUMA E DER ANSCHLUSSVARIANTEN ca 85 3 35 Abb 11 2 Ausf hrung mit Bajonettstecker...
Page 54: ...C5 AUSF HRUNGEN UND BESTELLNUMMERN 26...
Page 55: ...C5 ENGLISH DEUTSCH FRAN AIS Notice de montage...
Page 80: ...C5 VERSIONS ET NUM ROS DE COMMANDE 26...
Page 81: ...C5 ENGLISH DEUTSCH FRAN AIS...
Page 83: ...3 C5 1 05 6 C5 789 S S S S S S S A...
Page 84: ...C5 4 BC DE F GH G IJ C5 KLM 0 1 1 NO1 2 S 4 S S 5...
Page 85: ...5 C5 2 2 1 PQ 6 R S 0 9 7 TU 8 7 V TWX Y u 9...
Page 87: ...7 C5 4 TEDS TEDS Transducer Electronic Data Sheet 4 TEDS 9 17 B 5 L 6...
Page 88: ...C5 8 4 C 12 21 B...
Page 89: ...9 C5 5 5 1 D DMS OE 7 4 H 5 2 _ a 5...
Page 90: ...C5 10 6 6 F G 6 1 bcde H I Q L J R A 6 2 fg hi J K DIN EN 60529 IP 54 U 6 3 j L L...
Page 91: ...11 C5 7 7 1 kl V A S S S HBM EEK M S S 01 2 3 4 54 671 0 348 9 7 2 m n L 2 S 2 S 2 40 HRC S S...
Page 92: ...C5 12 Fex Fnom Fq Mb Fnom Fex Mb Fq C5 C5 N N K 40 HRC O...
Page 93: ...13 C5 7 3 o C5 6 R L R N N 2 C5 HBK ISO 376...
Page 96: ...C5 16 S 6 S S...
Page 99: ...19 C5 4 11 65 2 56 511 1 DDEFE GH...
Page 100: ...C5 4 20 85 3 35 511 2 DF FE GH...
Page 105: ...25 C5...
Page 106: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A05804 01 YC0 00 7 0111 0019...