HB Products HBPH-MK2 Instruction Manual Download Page 6

 

  

 

Instructional manual- HBPH-MK2 

– UK-004 

  

  

Page 6 

 

1.

 

Disconnect the red wire at the instrument or power supply and connect it to the multimeter (-) 
black.  

2.

 

Connect the multimeter (+) red to the instrument or power supply red wire output terminal.  

3.

 

Rinse the probe and place it in 7 pH buffer. Allow the temperature of the buffer and probe to 
stabilize at room temperature.  

4.

 

Check the offset of the probe by reading the multimeter.  The reading should be between 11 and 
13 mA.  If not, the probe is defective, or the salt bridge needs to be replaced.  If the offset is OK, 
please note the exact reading and proceed to the next step.  

5.

 

Rinse the probe and place it in 4 pH or 10 pH buffer.  Allow the temperature of the probe and 
buffer to stabilize at room temperature.  Now check the span of the probe by reading the 
multimeter.  If the probe is in 4 pH buffer, the reading should be between 2.75 and 3.5 lower than 
the reading obtained in 7 pH buffer.  

6.

 

If the probe is in 10 pH buffer, the reading should be between 2.75 and 3.5 higher than the reading 
obtained in 7 pH buffer.  

7.

 

If this test is not satisfied the probe is defective, or the salt bridge needs to be replaced. 

8.

 

If the span of the probe drops below 2.75mA than the probe still can be used. Adjustments will 
have to be made to the receiving device to compensate for the low span.  

 

 

 

Probe Cleaning  

The probe should be kept reasonably clean to avoid measurement errors.  Frequency of cleaning can only 

be determined by experience.  

To

 clean proceed as follows:  

1.

 

Rinse with clean warm water.  

2.

 

Soak the end of the probe in warm water and dish detergent for 3 or 4 minutes.  

3.

 

Brush the end of the probe, particularly the three electrodes with a soft bristle brush such as a 
toothbrush. Take care not to scratch the glass electrode.  

4.

 

If the probe is still not clean, it may have to be cleaned with acid. 

CAUTION: Do not acid clean 

probes used in processes containing cyanide solutions.

  Some experimentation may be required to 

determine the most suitable acid for your process.  Use the most dilute acid, which is effective. 
Normally 10 parts of water to one-part muriatic acid (HCl) is sufficient.  

Do not use hydrofluoric 

acid.

  

5.

 

Soak the probe for not more than 5 minutes in the chosen acid; then rinse thoroughly with clean 
warm water and soak in water for 3-5 minutes.  

6.

 

Calibrate the system in accordance with the instrument instruction manual.  

  

Summary of Contents for HBPH-MK2

Page 1: ...on 3 Cold Liquid applications 3 Electrical connection 4 Storage 4 Maintenance and calibration 4 Calibrating the Sensor 5 Checking the Sensor 5 Probe Cleaning 6 Replacement of Salt Bridge 7 Troubleshoo...

Page 2: ...period thoroughly clean the sensor put a piece of cotton ball with a few drops of water into the protective cap and place it on the sensor This keeps the junction from drying out which causes slow re...

Page 3: ...duit Do not run the power cable or control cables in the same conduit with the probe interconnect cable 3 Remove the protective plastic cap from the end of the probe before placing in service 4 The se...

Page 4: ...measurement If you use the sensor for a precise measurement of the pH value you need frequent checks of the calibration The drift of the sensor is maximum 0 03pH per day but that is the maximum figure...

Page 5: ...Allow the temperature of the buffer and probe to stabilize at room temperature 4 Write down the reading on the multimeter as mApH7 in the example called mApH7 The reading should be between 11 and 13 m...

Page 6: ...ied the probe is defective or the salt bridge needs to be replaced 8 If the span of the probe drops below 2 75mA than the probe still can be used Adjustments will have to be made to the receiving devi...

Page 7: ...th pH 7 solution or distilled water 5 Refill the chamber with a reference solution supplied with the salt bridge up to the tip of the electrode inside the chamber DO NOT OVERFILL It is important to le...

Page 8: ...iled Sensor is non repairable The output is outside of 4 20mA The sensor electronic has failed Sensor is non repairable Parts and Accessories Description Part 3 pack salt bridge replacement kit HBPH S...

Reviews: