![Hawk Hawk-i1 User Manual Download Page 55](http://html1.mh-extra.com/html/hawk/hawk-i1/hawk-i1_user-manual_2109380055.webp)
51
because they are not easy to be replaced. Battery replacement steps are listed as follows:
(1)
Unscrew the screws at the bottom of machine; remove the battery cover.
(2)
Unplug the battery cable and take out the battery.
(3)
Install the new battery. Please make sure the battery cable won’t be squeezed by the battery Cover.
Then install battery cover. After replacing new battery, please check its working condition.
(4)
The used battery can be sent back to manufacturer or distributor, or can be scrapped according to
legally proper way.
11.4 Normal repair procedures
The repair job should be performed by supplier or distributor. It needs to make a complete inspection
on machine after maintenance. If necessary
,
our company can offer circuit diagram and components
list to authorized maintenance personnel.
11.5 Maintenance for long-time storage
If the Infusion Pump will not be used for long time, it should be placed in packing carton and avoid
direct sunlight and keep it in cool and dry place. Refer to 12.2 for detailed storage conditions.
When using an Infusion Pump of long time storage, please refer to following steps before use:
(1)
Calibrate the Infusion Pump to ensure infusion accuracy and avoid possible medical accident.
(2)
Test Air Bubble and Occlusion alarm.
(3)
Test the working time and recharging time of battery to ensure the battery can still be used.
(4). Discharge and charge the battery every 3 months to ensure its service life.
11.6Recycling
The machines and its cable which have been used over its life span should be scrapped. For more
information, please contact manufacturer or our distributors. ( Whether it is used frequently or not and
whether it is repaired properly or not will impact infusion pump’s life span.)
1.The scrapped Infusion Pump and power cable can be sent back to manufacturer or distributor.
2. The used battery can be sent back to manufacturer or distributor, or can be scrapped
according to legally proper way.
12. Transport and storage
12.1 Precautions during transport
1. Place the product as per No. of layers indicated on packing carton.
2. Temperature: -20
℃
~45
℃
;