Model ‘11’ Single Syringe

55-1199

Model ‘11’ Dual Syringe

55-4154

Model ‘11’ 

Syringe Pump Series

User’s Manual

Summary of Contents for 11 series

Page 1: ...Model 11 Single Syringe 55 1199 Model 11 Dual Syringe 55 4154 Model 11 Syringe Pump Series User s Manual ...

Page 2: ...ding the Syringe 5 Fuse Replacement 6 Operation 7 Getting Started 7 Turn Pump ON 7 Function Keys and Run Indicator 7 Entering Syringe Size 8 Entering Flow Rate Range 8 Entering Flow Rate 8 Press Run 8 Check Syringe 8 Advanced Features 9 Volume Mode 9 Power Failure 9 Display Intensity 9 Changing Rates 10 Maintenance 10 Protecting Small Fragile Syringes 10 Appendices A Specifications 11 B Table of P...

Page 3: ...FITNESS FOR A PARTICULAR USE OR OF ANY OTHER NATURE Some states do not allow this limitation on an implied warranty so the above limitation may not apply to you If a defect arises within the one year warranty period promptly contact Harvard Apparatus Inc 84 October Hill Road Building 7 Holliston Massachusetts 01746 1388 using our toll free number 1 800 272 2775 Goods will not be accepted for retur...

Page 4: ...ine 16 character vacuum fluorescent display along with six membrane keys make this a most attractive but powerful easy to use syringe pump Only two entries required to start pumping syringe Inside Diameter mm and pumping flow rates The Flow rate can be changed while the Pump is running Two Modes of Operation Constant Flow Rate and Volume Dispense The new Pump 11 will operate continuously in RATE m...

Page 5: ...4 H a r v a r d A p p a r a t u s S y r i n g e P u m p M o d e l 1 1 Introduction New Pump 11 Front and Rear Views ...

Page 6: ...es on the display if power was disrupted since last use The procedure for loading syringes is the same for all models of the Pump 11 Model 55 1199 Single Syringe Pump Model 55 4154 Dual Syringe Pump Location Requirements for the Syringe Pump A sturdy level clean and dry surface Minimum of one inch 2cm clearance around the pump Adequate power supply Room temperature 4 C to 40 C 40 F to 104 F Relati...

Page 7: ...pen access door Remove fuse holder remove fuse s from holder Replace fuse s and fuse holder Replace holder with correct voltage marking is showing through access door window For continued fire protection replace fuse only with 250V fuse of the specified type and rating T 1 4 A 250V Fuse not replaceable by the operator is located on the power supply Fuse F 2 0A 250V Fuse Replacement WARNING USE IN ...

Page 8: ...AM Used to enter or query the syringe diameter SET This key is used to select which digit of the display is to be changed to move the decimal point and to move between modes Each time the set key is pressed the underline cursor below the digit or character on the display moves one step to the right It is used in conjunction with the ascending and descend ing keys When it the display shows the desi...

Page 9: ...the CONFIG mode Press the SET key again to move the underline cursor to the flow rate choices Choose your flow rate units while in the SET UNITS mode by pressing the or keys to scroll the four flow rate choices ml min µl min ml hr or µl hr Once the desired flow rate units are displayed press the ENTER key to return to the main SET CONFIG mode Press SET or ENTER again to put your desired flow rate ...

Page 10: ... volume dispense you can restart the pump at the place you stopped by pressing the RUN key again To exit the volume dispense mode set the target volume to 00 00 or turn off and on the pump via the main power switch NOTE In the event of a power failure the actual dispensed volume and the target volume are not retained in memory This means that while in volume dispense mode if a power failure occurr...

Page 11: ...until a new rate is entered Except for volume mode the flow rate can be changed while the pump is running As soon as the ENTER key is pressed the pump will change to the new flow rate To change rates from the keypad while in vol ume mode the pump must be stopped first 5 Maintenance No special maintenance is required other than keeping the pump 11 clean and dry Avoid liquid spills that may find the...

Page 12: ...ter vacuum fluorescent display and green run led Nonvolatile Memory Stores diameter rate and configuration settings Maximum Force 8 kg 17 lb Maximum Pressure 8 p s i with 50 60 cc syringe 100 p s i with 1 ml syringe Drive Motor 0 9 step angle motor Step Rate Minimum Maximum 1 pulse in 27 6 sec 200 steps sec Pusher Advance Step 0 8 µm pusher advance per motor step Pusher Travel Rate Minimum Maximum...

Page 13: ... Diameter 3 cc 8 95 mm 5 13 00 10 15 80 20 20 15 30 23 10 60 29 10 Sherwood Monoject Plastic Size Diameter 1 cc 4 65 mm 3 8 94 6 12 70 12 15 90 20 20 40 35 23 80 60 26 60 140 38 40 Popper Sons Inc Perfektum Glass Size Diameter 0 25 cc 3 45 mm 0 5 3 45 1 4 50 2 8 92 3 8 99 5 11 70 10 14 70 20 19 58 30 22 70 50 29 00 100 35 70 Appendix B Syringe Inside Diameter SGE Scientific Glass Engineering Size ...

Page 14: ...0001 0 0794 250 µl 2 30 0 7218 9 999 0 0121 197 0 0 0008 11 82 0 0001 0 1970 1 000 µl 3 26 1 451 9 999 0 0242 395 7 0 0015 23 75 0 0001 0 3959 1 ml 4 61 2 900 9 999 0 0484 791 8 0 0029 47 50 0 0001 0 7918 2 5 ml 7 28 7 232 9 999 0 1206 1 974 0 0073 118 4 0 0002 1 974 3 ml 8 66 10 24 9 999 0 1706 2 794 0 0103 167 6 0 0002 2 794 5 ml 10 30 14 50 9 999 0 2413 3 952 0 0145 237 1 0 0003 3 952 10 ml 14 ...