Device Functions Proline Promag 55
9 Block BASIC FUNCTION
Hauser
121
RECOVERY TIME ECC
(6442)
!
Note!
This function is not available unless the measuring device is equipped with the optional
electrode cleaning function (ECC).
Use this function to specify the recovery time for which the last flow value measured
prior to cleaning is retained. A recovery time is necessary as the signal outputs can
fluctuate after electrode cleaning on account of electrochemical interference voltages.
User input:
max. 3-digit number: 1 to 600 s
Factory setting:
60 s
"
Caution!
The last value measured prior to cleaning is output for the duration of the recovery time
(max. 600 s). This in turn means that the measuring system does not register changes in
flow, e.g. stoppage, during this time span.
CYCLE ECC
(6443)
!
Note!
This function is not available unless the measuring device is equipped with the optional
electrode cleaning function (ECC).
Use this function to specify the cleaning cycle for electrode cleaning.
User input:
Integer: 30 to 10080 min
Factory setting:
40 min
Function description
BASIC FUNCTION
➞
PROCESS PARAMETER
➞
ECC PARAMETER