Graphs include time tick marks that occur at hourly intervals for time scales 24
hours or less or at daily intervals for time scales greater than 1 day. Time tick
marks are labeled at intervals with the hour or day, labels being placed below
and to the right of tick marks. Tick marks with labels are drawn with short con-
tinuous lines, and tick marks without labels are drawn with dotted lines.
7.4
Alarm Menu
The ALARM menu allows the setting and viewing of alarms. The functions
that appear in this menu are ALARM SETTING, SENSOR ALARM, SYSTEM
ALARM, and ALARM VIEW. Alarm settings may be password protected (see
Section 7.5.4, Password).
The
ud
buttons are used to select the menu option. The
lr
buttons are
used to change the parameter setting. Press the
Enter
button to save the new
setting. Press the
Exit
button to cancel any changes made and exit to the menu.
7.4.1
Alarm Setting
The ALARM SETTING function is used to enable or disable visible and audi-
ble alarm notifications. The parameters are BEEP and DISPLAY.
When the BEEP parameter is ON, the thermo-hygrometer emits an audible
beep during an alarm condition. For the power alarm it is four short beeps ev-
ery 20 seconds. For the battery alarm it is three short beeps every minute.
For the sensor disconnect alarm it is two short beeps every measurement pe-
riod. For measurement alarms it is one long beep every measurement period.
60
Hart Scientific
7 Menu Functions
Figure 37
Alarm