![Hart Sceintific DewK 1620 User Manual Download Page 60](http://html1.mh-extra.com/html/hart-sceintific/dewk-1620/dewk-1620_user-manual_601476060.webp)
STATS PRINT, STATS RESET, and STATS CLEAR. The DAILY STATS can
be password-protected (see Section 7.5.4, Password).
The
ud
buttons are used to select the menu option. The
lr
buttons are
used to change the setting parameters. Press the
Enter
button to save the new
setting. Press the
Exit
button to cancel any changes and exit to the menu.
7.3.2.1
Stats Setting
The STATS SETTING function sets the daily statistics recording options. The
AUT REC enables (ON) or disables (OFF) automatic recording of the statistics.
The REC TIM may be set from 00:00 to 23:00. The REC TIM function sets the
hour just prior to which the statistics are automatically recorded when AUT
REC is ON. If REC TIM is 00:00, the statistics are recorded after the last mea-
surement of the day.
AUT RES enables (ON) or disables (OFF) automatic reset of the statistics and
may be set from 00:00 to 23:00. RES TIM sets the hour when the statistics are
automatically reset if AUT RES is ON. If RES TIM is 00:00, the statistics are
reset just prior to the first measurement of the day.
For example, to automatically store statistics for each day for measurements
made in a laboratory only between 8:00 in the morning and 5:00 (17:00) in the
afternoon, set the AUT REC to ON, REC TIM to 17:00, AUT RES to ON, and
RES TIM to 08:00. Note that statistics can also be manually reset using the
STATS RESET function (see Section 7.3.2.4).
DO NOT
manually reset the sta-
tistics if automatic recording is desired. The DAILY STATS settings and
STATS RESET can be password-protected to prevent inadvertent resetting of
52
Hart Scientific
7 Menu Functions
Figure 28
Daily Stats