8.4.6.4
CALibrate<chn>:DATE:CALibrate (<year>,<month>,<day>)
Sets the sensor calibration date for the specified channel.
Example: CAL1:DATE:CAL 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <year>,
<month>, and <day> parameters must all be specified. The <year> parameter is
a four-digit number. The <month> parameter is a one or two-digit number (1 to
12). The <day> parameter is a one or two-digit number (1 to 31). This com-
mand is password protected (see Section 8.4.10, Password Commands).
8.4.6.5
CALibrate<chn>:DATE:DUE?
Returns the sensor calibration due date for the specified channel.
Example: CAL1:DATE:DUE?
Response: 2004,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is re-
turned in the format, <year>,<month>,<day>.
8.4.6.6
CALibrate<chn>:DATE:DUE (<year>,<month>,<day>)
Sets the sensor calibration due date for the specified channel.
Example: CAL1:DATE:DUE 2004,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <year>,
<month>, and <day> parameters must all be specified. The <year> parameter is
a four-digit number. The <month> parameter is a one or two-digit number (1 to
12). The <day> parameter is a one or two-digit number (1 to 31). This com-
mand is password protected (see Section 8.4.10, Password Commands).
8.4.6.7
CALibrate<chn>:EXPired?
Returns the sensor calibration expiration state for the specified channel.
Example: CAL1:EXP?
Response: 0
The CALibrate suffix, <chn>, specifies the channel (1 or 2). A value of 1 is re-
turned if the sensor calibration has expired. A value of 0 is returned if the sen-
sor calibration has not expired.
8.4.6.8
CALibrate<chn>:PARameter:OFFSet<num>?
Returns the sensor calibration offset parameter for the specified channel.
Example: CAL1:PAR:OFFS2?
1620 the “DewK”
113
8 Digital Communications Interface