8.4.1.33
ALARm:TEMPurature<chn>:LOWer:ENABle?
Returns the state of the lower temperature alarm enable for the specified
channel.
Example: ALAR:TEMP1:LOW:ENAB?
Response: 0
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if the lower temperature alarm is enabled. A value of 0 is returned if
the lower temperature alarm is disabled.
8.4.1.34
ALARm:TEMPurature<chn>:LOWer:ENABle <bool>
Sets the state of the lower temperature alarm enable for the specified channel.
Example: ALAR:TEMP1:LOW:ENAB 1
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The <bool> pa-
rameter turns the lower temperature alarm on (1 or ON) or off (0 or OFF). This
command may be password protected (see Section 8.4.10, Password
Commands).
8.4.1.35
ALARm:TEMPurature<chn>:LOWer:LIMit? [MIN|MAX|DEF]
Returns the lower temperature alarm limit for the specified channel in degrees
C or F.
Example: ALAR:TEMP1:LOW:LIM?
Response: 18.00
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The value re-
turned is always in the current temperature units (C or F). Specifying the MIN,
MAX, or DEF parameter returns the minimum, maximum, or default value
respectively.
8.4.1.36
ALARm:TEMPurature<chn>:LOWer:LIMit <float>|MIN|MAX|DEF
Sets the lower temperature alarm limit for the specified channel in degrees C or
F.
Example: ALAR:TEMP1:LOW:LIM 20.00
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The <float> pa-
rameter specifies the value for the lower temperature alarm limit. Specifying
the MIN, MAX, or DEF parameter sets the lower temperature alarm limit to the
minimum, maximum, or default value respectively. The <float> value is always
in the current temperature units (C or F). This command may be password pro-
tected (see Section 8.4.10, Password Commands).
1620 the “DewK”
93
8 Digital Communications Interface