Specifications
7
•
•
•
•
•
•
Chapter 2
• • • • • •
Specifications
Chapter 2 provides specifications for HI 6500 series instruments. The specifications
listed are designed to assist in the installation, operation and troubleshooting of your
instrument. All service personnel should be familiar with this section before installing
or repairing the instrument.
Basic Specifications
General
Number of Channels
•
1 Channel per unit
Update Rate
•
HI 6500 and HI 6510:
110 times per second (processed weight, display,
communications)
•
HI 6500-XP and HI 6510-XP:
350 times per second (processed weight, display,
communications)
•
Unprocessed weight (A/D conversion) 4800 times per second
Resolution
•
HI 6500 and HI 6510:
Displayed - 1:10,000
•
HI 6500-XP and HI 6510-XP:
Displayed
1:30,000
•
Addressable resolution - 1:999,999
WAVERSAVER
•
User Selectable
OFF
7.50 Hz
3.50 Hz
1.00 Hz (default)
0.50 Hz
0.25 Hz
WAVE (HI 6500-XP and HI 6510-XP Only)
•
Patent pending dynamic filtering improves static weight readings by a factor of 3X
Averages
•
1 to 255 User-selectable in Single Increments
Power
•
12-24 VDC
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......