106
Chapter
5
•
•
•
•
•
•
NOTE
All the samples must be the same weight, type and size.
Be sure the scale has been Zero’d before determining piece count.
Hint: When the weight of individual pieces varies, it is better to take the Weight Sample
of a larger number of pieces when the weight of individual pieces varies.
On the
Web page
: first select Count Enabled, and then either enter the weight of a unit in
the unit weight field, or use the weigh sample field entering the correct number of pieces in
the Sample size field as shown above. Then click on the weigh sample field.
Security Parameter Menu
A detailed description of the security features is provided in Chapter 8.
•
Calibration Lock
o Calibration Lock
o Calibration Password
•
Configuration Lock
o Configuration Lock
o Password
•
Display Lock
o Display Lock
o Display Password
•
Keypad Lock
o Keypad Lock
o Keypad Password
•
Read Only Lock
o Read Only Lock
o Read Only Password
NOTE
The security parameters can be accessed via the front panel display by entering a
preset password, or directly through the network connection. There is no way to
bypass the security password. We recommend you keep your passwords in a secure
location.
The security menu allows the user to lock out different levels of the menu hierarchy.
The user-configurable security settings only limit access through the front panel to
ensure the consistency of the instrument setup and weighing process. Parameter
configuration through the network is unaffected by these security settings.
The factory default for the security menu is to disable all lock modes and to preset each of
the security passwords to 1234. Below is another example where each password in the
security menu is different.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......