Chapter 4 - Setup
4-13
The configuration data is sent from the PLC to the HI
1769-WS module at power-up. The module uses these
parameters provided that:
1.
The parameters are in the correct range. Ille-
gal values will be rejected.
2.
The “CopyConfig” word (0 for channel 0, 24
for channel 1) is set to 1.
•
INT parameters are 2 byte integers
•
DINT parameters are 4 byte integers
•
REAL parameters are 4 byte IEEE
floating point numbers
NOTE:
Parameters labeled “REAL or DINT” will be inter-
preted as floating point or integer according to the
value of the “METRIC” parameter of the channel. If
bit 6 (0x40) of METRIC is set the parameter is float-
ing point. If bit 6 is not set, it is a fixed point integer,
with 0-7 decimal places as determined the first 3 bits
of the METRIC parameter.
Parameter
Offset (In
Words)
Data Type
Ch0CopyConfig
0
INT
Ch0ChanActive
1
INT
Ch0Metric
2
INT
Ch0Waversaver
3
INT
Ch0NumAverages
4
INT
Ch0ZeroTrackEnable
5
INT
Ch0AutoZeroTolerance
6
REAL or INT
Ch0MotionTolerance
8
REAL or INT
Ch0ZeroTolerance
10
REAL or INT
Ch0SpanWeight
12
REAL or INT
Summary of Contents for HI 1769-WS
Page 14: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 1 6...
Page 27: ...Chapter 3 Installation 3 9 FIG 3 7 RAIL FASTENERS IN THE CLOSED POSITION...
Page 60: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 4 26...
Page 74: ...HI 1769 WS HI 1769 2WS MANUAL 6 8...
Page 82: ...HI 1769 WS HI 1769 2WS MANUAL Weighing System Tests 1 4 wrist strap grounding device 3 2...