HI 1746-WS WEIGH SCALE MODULE
4-2
FIG. 4-1 MODULE LEDS
LEDS
Run/Fault LED
Flashing Green Self-Test is running.
Flashing Green (5Hz Flash Speed) - Self-Test
Failed, or no calibration.
Steady Green
Self-Test Passed
Flashing Red
Weigh Scale Module has a fault.
Steady Red
Module has a major problem.*
NOTE:
* Contact Hardy Customer Support for assistance.
Comm. LED
Steady Green
Normal Operation
Flashing Green
Block transfer is being performed.
LED is Off
No Communication
C2 LED
Steady Green
A C2 load cell(s) has been found.
LED is Off
No C2 load cells have been found.
Steady Red
Unable to read loadcell data or the
loadcell data is in error.
Setting Up
Communications
Between the SLC and
the HI 1746-WS Weigh
Scale Module
To set up communication between the SLC 5XX and
the HI 1746-WS Weigh Scale Module you will need
to select or enter the following information into the I/
O Configuration section of your software:
Summary of Contents for C Series
Page 8: ...HI 1746 WS WEIGHT SCALE MODULE vi...
Page 10: ...HI 1746 WS WEIGH SCALE MODULE II...
Page 46: ...HI 1746 WS WEIGH SCALE MODULE 4 16 I O Setup Ladder Logic for Remote Mode Operation...
Page 47: ...Chapter 4 Setup 4 17...
Page 54: ...HI 1746 WS WEIGH SCALE MODULE 4 24 SETPARAMS Remote Mode of Operation...
Page 55: ...Chapter 4 Setup 4 25...
Page 56: ...HI 1746 WS WEIGH SCALE MODULE 4 26...
Page 60: ...HI 1746 WS WEIGH SCALE MODULE 4 30...
Page 69: ...Chapter 5 Calibration 5 9...
Page 70: ...HI 1746 WS WEIGH SCALE MODULE 5 10 FIG 5 5 C2 LADDER LOGIC EXAMPLE LOCAL MODE OF OPERATION...
Page 74: ...HI 1746 WS WEIGH SCALE MODULE 5 14...
Page 84: ...HI 1746 WS WEIGH SCALE MODULE 6 10...
Page 92: ...HI 1756 WS MANUAL...