background image

 

HA1712-UM 

Rev NI 

April 30, 2020 

 

User Manual 

 

Page 11 of 12

 

INSPECTION 

Daily Inspection 

1.  Visually inspect the outer pouch for obvious damage (e.g. puncture, tear, swelling, etc.).  If the outer 

pouch is obviously damaged, it will be necessary to open the outer pouch and conduct further 
inspection of the inner packaged contents (See 

Inspection 

below). 

2.  Ensure the red inflator lanyard extends out of the top flap of the pouch. 

After Any Inflation Event 

1.  After each field-use inflation event, the MD3080s are to be returned to Harbour Air for inspection 

and recertification. 

Annually 

In addition to the inspection specified before each use, perform the following at least every year or 
more often, if exposed to potential damage or used in extreme temperature conditions.  

1.  Open the top flap of the pouch, remove the inner package from the pouch.   
2.  Visually confirm that the anti-tamper seal that crosses the inflator lever arm is present and secure.  

It is not necessary to open the inner package.  The presence of the anti-tamper seal indicates that 
the inflator’s lever arm has not moved since the device was last armed and that it is ready for use.   

3.  If the anti-tamper seal is dislodged or missing, return the IFD to Harbour Air for inspection and 

recertification prior to further use. 

4.  Inspect the inner pouch to ensure it is not torn or punctured.  If the inner pouch is damaged, return 

the IFD to Harbour Air for inspection and recertification prior to further flight. 

5.  If there is no obvious damage to the inner pouch and the anti-tamper seal is secured across the 

lever arm, the device is ready for service. 

6.  Visually inspect the outer pouch for obvious damage (e.g. puncture or tear).  If outer pouch is 

damaged, replace with new outer pouch.  The sealed inner pouch and the outer pouch can be 
separated for replacement by detaching the snap from the d-ring shown in Figure 8. 

7.  Prior to replacing the inner pouch into the outer pouch ensure that the snap assembly secures the 

lanyard through the d-ring. 

8.  Ensure that the lanyards are secured by the hook and loop at the finger pull loop (Figure 6). 
9.  Insert the inspected inner pouch in the outer pouch such that the inflator handle extends outside the 

flap, as shown below: 

 

Figure 9 

– Inflator Handle Position 

Summary of Contents for HA1712-UM

Page 1: ...l Flotation Device IFD User Manual HA1712 UM Rev NI April 30 2020 Warning This document is for the use of licensed users only No part of this document may be reproduced transmitted or transcribed in any form or by any means without the express permission of Harbour Air Seaplanes ...

Page 2: ...UCTION 4 DESCRIPTION 4 INSTALLATION INSTRUCTIONS CONSIDERATIONS 6 APPROVAL 6 WEARING THE IFD 7 INFLATING THE IFD 7 MANUAL INFLATION 7 ORAL INFLATION 7 SPECIAL CONSIDERATIONS 8 OPERATING THE IFD IN THE WATER 9 MAINTENANCE 10 CLEANING OF THE IFD 10 STORAGE 10 INSPECTION 11 DAILY INSPECTION 11 AFTER ANY INFLATION EVENT 11 ANNUALLY 11 5 YEAR INSPECTION 12 WARRANTY 12 CONTACT 12 ...

Page 3: ...ision Description of Revisions Affected Pages Date NI New Issue All April 30 2020 LIST OF FIGURES Figure 1 MD3080 Harbour Air Individual Flotation Device 4 Figure 2 MD3080 Comprising of outer pouch MA3078 and sealed inner pouch with bladder MA3079 5 Figure 3 Inflated Bladder MA3079 5 Figure 4 Waist Belt Extender MA3077 6 Figure 5 IFD Label 6 Figure 6 Pull finger loop to separate IFD bladder from o...

Page 4: ...y to be used within an overwater aviation environment Figure 1 Figure 1 MD3080 Harbour Air Individual Flotation Device The MD3080 is comprised of the MA3078 outer pouch Figure 2 and MA3079 bladder Figure 3 There is also an optional waist belt extender MA3077 that can be ordered and stowed separately when not in use Figure 4 WARNING Do NOT inflate the IFD while inside an enclosed cabin space as tha...

Page 5: ...2 Figure 2 MD3080 Comprising of outer pouch MA3078 and sealed inner pouch with bladder MA3079 Figure 3 Inflated Bladder MA3079 Lanyard Buckle Inflator Inflator Handle Outer Pouch Inner Pouch Anti Tamper Seal Bladder Oral Inflator Whistle Arm Straps Handle ...

Page 6: ...manual the Harbour Air MD3080 Individual Flotation Device IFD can greatly increase a user s chance of survival in the water Installation Instructions Considerations There are no installation instructions for the Harbour Air IFD Instructions for donning and usage are given in the following sections The IFD is intended to be worn around the user s waist and thus it is not necessary to install into a...

Page 7: ...ing the inflator releases CO2 gas into the inflation cell rapidly inflating the IFD The IFD should inflate within five seconds If the IFD does not completely exit its innermost plastic pouch this pouch can be torn by hand along the edge of the pouch s seal to aid removal of the IFD bladder WARNING Do NOT inflate the IFD while inside an enclosed cabin space as that will prevent or severely inhibit ...

Page 8: ... inflation To compensate use the oral inflator tube to top up the IFD NOTE When temperatures are extremely cold below 30 C the IFD must be inflated while in the water not in the open air WARNING Do NOT inflate the IFD while inside an enclosed cabin space as that will prevent or severely inhibit emergency egress Should there be a need to separate the bladder from the outer pouch due to snag entrapm...

Page 9: ...adder While the IFD bladder is reversible with arm straps located on both the top and bottom the oral inflation tube is located only on the top of the bladder Ideally once in the water you are facing towards the oral inflation tube and the safety whistle should they be needed Figure 7 It may be necessary to jump into the water with the device either stowed deflated or inflated maximum height of 3 ...

Page 10: ...py water Rinse the outer pouch with clean water Hang the outer pouch to dry on a plastic coat hanger To clean the inner pouch Sponge down the inner pouch exterior with a wet soapy cloth Do not saturate the edge of the inner pouch above the heat seal line as this can allow unwanted moisture to penetrate the IFD bladder packaging Hand wash or sponge down the exposed lanyard with warm soapy water Han...

Page 11: ...indicates that the inflator s lever arm has not moved since the device was last armed and that it is ready for use 3 If the anti tamper seal is dislodged or missing return the IFD to Harbour Air for inspection and recertification prior to further use 4 Inspect the inner pouch to ensure it is not torn or punctured If the inner pouch is damaged return the IFD to Harbour Air for inspection and recert...

Page 12: ... or if it is suffering wear and tear In the case that a product does have a manufacturing defect deemed by Harbour Air and is repairable Harbour Air may choose to not replace but to repair the product free of charge The decision on whether to repair or replace a product is solely up to the discretion of Harbour Air The warranty applies to the original purchaser only and proof of purchase is requir...

Reviews: