User Manual
43
3. Use Utilities ->Setup ->Calibration to calibrate.
4. Follow the prompts to connect the multimeter, input the multimeter reading into the
"Input DMM measurement value", press the Enter key to confirm, and proceed to the
next step.
1
)
Channel: When the calibration channel changes, the corresponding channel wiring
needs to be changed.
2
)
Sense: 2-Wire/4-Wire mode. If the output is before calibration, all channels need to
be set to 2-Wire mode before calibration; If the output is calibrated, all channels need to
be set to 4-Wire mode before calibration. Source settings>2W/4W Set 2-Wire/4-Wire
mode.
3
)
Type: Current calibration current/voltage type. When V
→
A or A
→
V, pay attention
to changing the wiring.
4
)
Unit: The unit of input data in the DMM measurement value.
5. After calibration, the power automatically exits the calibration interface and can be used
normally after restarting.
Attention: All channels need to be calibrated at once to save the calibration results;
After self calibration by the user, the factory calibration data cannot be restored;
During the calibration process, if input data is incorrect, it can lead to inaccurate power
output after calibration. Exit and recalibrate at this time.
11.4 Error message
Error displays the error information of the instrument. The error information list is arranged
in the order in which it was generated, that is, the most recent error is displayed at the top.
Press "Next" to view the next page.
11.5 Capture
Capture can set the properties of the screenshot file, including the save path and file name.
The default picture format is .bmp.