27
6.3 Parallel
In parallel mode, the channels are internally connected in parallel to form a single output.
The channel 2 window of the main interface displays "PARALLEL with CH1". Channel 1 is
the master and channel 2 is the slave. The "+" and "-" binding posts of channel 1 are the
two output terminals of the parallel circuit. The output current is twice the set current of
channel 1, and the output voltage is the set voltage of channel 1. The external wiring
method of parallel mode is as follows:
6.4 Tracking
In tracking mode, channel 1 and channel 2 are associated with each other, and the
voltages track each other. Changing the set voltage of either channel will change the set
voltage of the other channel accordingly; the current of the two channels will not be
affected.
Note: When using the tracking function, make sure that the OCP setting of each channel
does not affect the normal output of the voltage.
Tracking mode can also be set by long pressing "Tracking" on the front panel, and short
pressing "Tracking" to exit. After setting, the “Tracking” logo will be displayed between
channel 1 and channel 2 on the main interface.
8. Output list
The output list can be output according to the conditions set by the list. Output conditions
include trigger point, output duration, output cycle times, etc.