background image

The "four-in-one" portable meter

WATER TEST

GENERAL DESCRIPTION

The HANNA Water Test is a "four-in-one" portable meter designed for
fast and accurate analysis and monitoring of water in biological
applications, industrial wastewater, rivers, lakes, etc. It measures the
main parameters in water analysis, namely, pH,  oxidation-reduction
potential 
(REDOX), conductivity and temperature.  All is needed for
the analysis is a small sample of the water which is poured into the
compartment and the 4 measurements are immediately available.
The instrument has a complete pH range from 0 to 14, an ORP range
from -1000 to +1000 mV, a conductivity range from 0 to 2000 µS/cm
and a temperature range from 0 to 60°C. The instrument can be
calibrated for pH and conductivity  and comes with replaceable batteries.

SPECIFICATIONS

pH

ORP

Conductivity

Temperature

RANGE

0.0 to 14.0 pH

±1000 mV

0 to 1999 µS/cm

0.0 to 60.0°C

RESOLUTION

0.1 pH

1 mV

1 µS/cm

0.1°C

ACCURACY

±0.2 pH

±5 mV

±2% Full Scale

±1°C

(@20°C/68°F)

TYPICAL EMC

±0.4 pH

±2mV

±2% Full Scale

±1°C

DEVIATION

CALIBRATION

pH:

Manual double point through Offset and Slope trimmers

µS/cm: Manual single point through trimmer

BATTERY

3 x 1.4V batteries for approx. 200 hours of continuous use

DIMENSIONS

190 x 85 x 85 mm
(7.4 x 3.3 x 3.3")

WEIGHT

260 g (9.2 oz.)

Stay in Touch with the Manufacturer !

DECLARATION OF CONFORMITY

We

Hanna Instruments Srl
V.le delle industrie 12
35010 Ronchi di Villafranca (PD)
ITALY

herewith certify that the pH / ORP / Conductivity / Temperature meter

has been tested and found to be in compliance with the following regulations:

IEC 801-2

Electrostatic Discharge

IEC 801-3

RF Radiated

EN 55022

Radiated, Class B

Date of Issue: 12-12-1995

D.Volpato - Engineering Manager

On behalf of

Hanna Instruments S.r.l.

Water Test

MAINTENANCE

• Separate the compartment from the tester.

• Remove deposits formed on the liquid  junction

and the pH glass bulb by  soaking in HI 7061
solution for 1½ hour.

• Soak the platinum wire in acetone for half hour.

• Rinse the bulb, platinum wire and liquid junction

thoroughly with distilled water

• Reassemble the compartment and the tester.

• Fill the compartment with tap water.

BATTERY REPLACEMENT

When the Water Test cannot be switched on or the
display fades, pull out the battery cover and replace
all three 1.4V batteries, paying attention to their po-
larity. Battery replacement must only take place in a
non hazardous area using the battery types speci-
fied in this instruction manual.

ACCESSORIES

HI 7004M

Buffer solution pH 4.01 (230 ml)

HI 7004L

Buffer solution pH 4.01 (460 ml)

HI 7007M

Buffer solution pH 7.01 (230 ml)

HI 7007L

Buffer solution pH 7.01 (460 ml)

HI 7010M

Buffer solution pH 10.01 (230 ml)

HI 7010L

Buffer solution pH 10.01 (460 ml)

HI 7031M

Conductivity solution 1413 µS/cm @25°C (230 ml)

HI 7031L

Conductivity solution 1413 µS/cm @25°C (460 ml)

HI 7061M

General Cleaning solution (230 ml)

HI 7061L

General Cleaning solution (460 ml)

BATT009/P

1.4V battery (100 pcs)

HI 76501/P

Calibration screwdriver (20 pcs)

RECOMMENDATIONS FOR USERS

Before using this product, make sure that it is entirely suitable for
the environment in which it is used.

Operation of this instrument in residential area could cause
unacceptable interferences to radio and TV equipments,
requiring the operator to take all necessary steps to correct
interferences.

The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharge.

Avoid touching this glass bulb at all times.

For calibration it is recommended to use an antistatic
screwdriver.

Any variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance.

To avoid electrical shock, do not use this instrument when voltages
at the measurement surface exceed 24 VAC or 60 VDC.

To avoid damages or burns, do not perform any measurement in
microwave ovens.

SWATERR1

03/96

ON / OFF

SWITCH

RANGE SWITCH

LIQUID

CRYSTAL

DISPLAY

MAXIMUM

IMMERSION

LEVEL

BLACK

COMPARTMENT

FRONT

BATTERY

COVER

OFFSET & SLOPE

CALIBRATION

TRIMMERS

COVER

BACK

Water Test

ON / OFF

4   pH  7

RANGE

IMMERSION LEVEL

Water Test

ON / OFF

4   pH  7

RANGE

BATTERY

COVER

IMMERSION LEVEL

Reviews: