background image

Troubleshooting

• In calibration mode, the “--- Err” message is displayed if the reading is 

outside accepted range or buffer solution is not recognized. 

• If the probe is not in buffer solution, “pH 14.00” or “pH 0.00” is displayed 

blinking.

• If measured temperature is outside accepted range, “0 

°

C” or “50 

°

C” is 

displayed blinking.

Care & Maintenance

Please read the information below to ensure highest possible accuracy.
• Fresh buffers should be used for each calibration, once the sachets are open 

the buffer value can change over time.

• If the measurements are taken sucessively, rinse the probe thoroughly in 

purified water to eliminate cross-contamination.

• When not in use, add a few drops of storage solution to the protective cap to 

keep the junction hydrated. If storage solution is not available, pH 4.01 or pH 
7.01 buffer can be used.

• For improved accuracy, a two-point calibration is recommended.
• When dirty, the junction can be pulled out to expose a fresh section. Pull 

out 3 mm (1/8”) of the cloth, making sure that the reference compartment 
remains covered. This procedure can be repeated approximately 9 times, 
renewing the tester’s life.

• If the electrode or junction is dirty, soak the tip in 

HI7061

 General purpose 

cleaning solution for 30 minutes. Rinse with distilled water and follow 
cleaning procedure.

Film deposits and other impurities that accumulate with use must be removed 
periodically off the membrane or junction.
• 

General:

 Soak in 

HI7061

 General cleaning solution for approximately 1 hour.

• 

Protein:

 Soak in 

HI7073

 Protein cleaning solution for 15 minutes.

• 

Inorganic:

 Soak in 

HI7074

 Inorganic cleaning solution for 15 minutes.

• 

Oil & grease:

 Rinse with 

HI7077

 Oil and grease cleaning solution.

Note: 

After performing any of the cleaning procedures rinse the electrode 

thoroughly with distilled water, and soak in 

HI70300

 Storage solution for 

at least 1 hour before taking measurements.

Battery Replacement

Battery life percentage indicator is displayed at power on. If the battery 
level drops below 10%, the battery symbol starts blinking. If the battery is 
discharged, “dEAd bAtt” is displayed and then the tester turns off.

08

08

To change the CR2032 Lithium-ion battery, turn the battery cover located on the 
back of the tester counterclockwise to unlock. Remove the cover and replace the 
battery with positive (+) side facing out.

Note: 

Only use the battery type specified in the manual. Old batteries 

should be disposed in accordance with local regulations.

Accessories

pH Buffer Solution

HI70004P

pH 4.01 buffer solution, 20 mL sachet (25 pcs.)

HI70007P

pH 7.01 buffer solution, 20 mL sachet (25 pcs.)

HI70010P

pH 10.01 buffer solution, 20 mL sachet (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)

HI770710P

pH 10.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)

Electrode Cleaning Solution

HI7061M

General purpose cleaning solution, 230 mL

HI7073

Protein cleaning solution

HI7074

Inorganic cleaning solution

HI7077

Oil and Grease cleaning solution

Electrode Storage Solution

HI70300M

Electrode storage solution, 230 mL

Certification

All Hanna Instruments conform to the 

CE European Directives.

Disposal of Electrical & Electronic Equipment.

 The product 

should not be treated as household waste. Instead hand it over 
to the appropriate collection point for the recycling of electrical 
and electronic equipment which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do 

not dispose of them with other household waste. Hand them over 
to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative 
consequences for the environment and human health. For more information, 
contact your city, your local household waste disposal service, the place of 
purchase or go to www.hannainst.com.

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific 
application and for the environment in which it is used. Avoid touching the 
electrode area. Any variation introduced by the user to the supplied equipment 
may degrade the tester’s performance. For yours and the tester’s safety do not 
use or store the tester in hazardous environments.

Warranty

This tester is warranted for a period of one year against defects in workmanship 
and materials when used for their intended purpose and maintained according 
to instructions. This warranty is limited to repair or replacement free of charge. 
Damage due to accidents, misuse, tampering or lack of prescribed maintenance 
is not covered. If service is required, contact your local Hanna Instruments 
Office. If under warranty, report the model number, date of purchase, serial 
number and the nature of the problem.
If the repair is not covered by the warranty, you will be notified of the charges 
incurred. If the instrument is to be returned to Hanna Instruments, first obtain 
a Returned Goods Authorization (RGA) number from the Technical Service 
department and then send it with shipping costs prepaid. When shipping any 
instrument, make sure it is properly packaged for complete protection.

Hanna Instruments reserves the right to modify the design, construction, or 

appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the written 

consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 

02895, USA.

IST98108N 12/19

RoHS

compliant

HI98108

Pocket-sized pH Tester

INSTRUCTION MANUAL

Summary of Contents for pHep+ HI98108

Page 1: ...HI98108 Pocket sized pH Tester INSTRUCTION MANUAL ...

Page 2: ... CR2032 3V Lithium ion battery Instrument quality certificate Instruction manual Note Save all packing material until you are sure that the tester works correctly Any damaged or defective item must be returned in its original packing material with the supplied accessories General Description Intended Use HI98108 pHep is a compact and waterproof pH tester that can be used in laboratories and indust...

Page 3: ...2032 3V Lithium ion included Battery life Approx 800 hours of continuous use Auto off 8 minutes 60 minutes or disabled Environment 0 to 50 C 32 to 122 F RH 100 max Dimensions 160 x 40 x 17 mm 6 3 x 1 6 x 0 7 Weight without battery 65 g 2 3 oz Functional Description LCD Display Stability tag Battery life percentage indicator Battery tag pH electrode Liquid Crystal Display CAL button ON OFF button R...

Page 4: ...nt mode To turn the tester off from measurement mode press ON OFF button OFF will be displayed on the second LCD line and the tester will turn off Calibration Mode To enter calibration mode press CAL button and the tester displays CAL message indicating that the instrument is ready for measurement Setup Setup menu allows the selection of temperature unit and auto off time The default settings are ...

Page 5: ...H 7 01 buffer Press CAL to save current calibration Stor will be displayed and the tester returns to measurement mode Two Point Calibration 1 Follow the first 2 steps from One Point Calibration using pH 7 01 buffer 2 After the first buffer is recognized place the tip of the electrode in pH 4 01 or pH 10 01 buffer 3 When the buffer is recognized the stability tag will disappear Press CAL button to ...

Page 6: ...curacy a two point calibration is recommended When dirty the junction can be pulled out to expose a fresh section Pull out 3 mm 1 8 of the cloth making sure that the reference compartment remains covered This procedure can be repeated approximately 9 times renewing the tester s life If the electrode or junction is dirty soak the tip in HI7061 General purpose cleaning solution for 30 minutes Rinse ...

Page 7: ... specified in the manual Old batteries should be disposed in accordance with local regulations Accessories pH Buffer Solution HI70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI70010P pH 10 01 buffer solution 20 mL sachet 25 pcs HI77400P pH 4 01 7 01 buffer solution 20 mL sachet 10 pcs 5 each HI770710P pH10 01 7 01buffersolution 20mLsachet 1...

Page 8: ...safety do not use or store the tester in hazardous environments Warranty This tester is warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered...

Reviews: