pH 502 & mV 602 Series

Panel-mounted,
Microprocessor-based
pH and ORP Controllers

Instruction Manual

Summary of Contents for mV 602 Series

Page 1: ...pH 502 mV 602 Series Panel mounted Microprocessor based pH and ORP Controllers Instruction Manual...

Page 2: ...ments All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Dear Customer Thank you for choosing a Hanna Product Please read this in...

Page 3: ...re that no damage has occurred during shipping If there is any noticeable damage notify your Dealer or the nearest Hanna Customer Service Center immediately Note Save all packing materials until you a...

Page 4: ...d reverts back to normal mode in idle or control phases with the measure ment on the display In pH 502 series during pH calibration alternately displays pH buffer value or current temperature 3 SETUP...

Page 5: ...alEMCDeviation 0 2 pH pH 502 series 10 mV mV 602 series 0 5 C Calibration pH automatic 1 2 or 3 point at pH 4 01 7 01 10 01 ORP automatic 2 point at 0 and 350 or 1900 mV Temperature Compensation pH 50...

Page 6: ...eys PowerSupply Connecta3 wirepowercable to the terminal strip while paying attention to the correct line L earth PE and neu tral N terminal connections Power 115 Vac 100 mA 230 Vac 50 mA Line Contact...

Page 7: ...no H1 0 to 4000 mV 50 mV 14 Relay 1 deviation 0 50 to 14 00 pH 1 pH no D1 25 to 4000 mV 50 mV 15 Relay 1 reset time 0 1 to 999 9 minutes 999 9mins no 16 Relay 1 rate time 0 0 to 999 9 minutes 0 0 mins...

Page 8: ...n S1 H1 S2 H2 S2 LA HA S1 If M1 2 and M2 1 then S2 H2 S1 H1 S1 LA HA S2 Code Valid Values Default PW 22 Relay 2 setpoint S2 0 00 to 14 00 pH 6 00 pH no 2000 to 2000 mV 500 mV 23 Relay2hysteresis H2 0...

Page 9: ...hen S1 S2 H2 S2 LA HA S1 D1 If M1 2 and M2 3 then S1 H1 S2 S1 LA HA S2 D2 If M1 4 and M2 1 then S1 S2 H2 S1 D1 LA HA S2 If M1 1 and M2 4 then S1 H1 S2 S2 D2 LA HA S1 If M1 3 and M2 4 then S1 S2 S2 D2...

Page 10: ...oint Hysteresis Setpoint t0 t0 Tc t0 2Tc t0 3Tc An upper boundary is imposed for acid base dosage time when relays are energized continuously i e when relay works in ON OFF mode or in PID mode but in...

Page 11: ...S2 the deviation D1 or D2 the reset time the rate time and the proportional control mode period Tc from 1 to 30 minutes for models with electromechanical relays and from 5 seconds to 30 minutes for mo...

Page 12: ...arm function operates in a Normally Closed state and hence alarm is triggered if the wires are tripped or when the power is down 1 Starting from a solution with a pH or mV value different from the dos...

Page 13: ...off the alarm in abnormal circumstances for example if the dos ing terminals are closed for too long a period In both cases the red LED s will also provide a visual warning signal The Fail Safe mode...

Page 14: ...e The type voltage or current and the range of the output analog signal is selectable through the jumpers on the power board Configurations of the switch are as follows Output Switch 1 Switch 2 Switch...

Page 15: ...on the disk s label Contact your Hanna Dealer to request a copy SPECIFICATIONS The RS485 standard is implemented in pH502 mV602 with the following characteristics Data rate up to 9600 bps Communicati...

Page 16: ...characters of a command It is possible to send commands to change the controller settings or to simply ask information on the controller status Following is the complete list of commands available Com...

Page 17: ...nnot answer the request e g the password was not sent the controller is in setup mode the setup item is not available in that model etc Note The controller answers to the GET command with the same dat...

Page 18: ...he calibration procedure Initial Preparation Pour small quantities of pH 7 01 HI 7007 and pH 4 01 HI 7004 and or pH 10 01 HI 7010 solutions into indi vidual beakers If possible use plastic beakers to...

Page 19: ...ue For best accuracy however it is recommended that a two point calibration is performed HI 7007 CALIBRATION HI 7007 RINSE For accurate calibration use two beakers for each buffer solution the first o...

Page 20: ...ration Two point Calibration Proceed as described above for one point cali bration usingpH 7 01asthefirstpoint butdo notquitcalibrationbypressingCALattheend Note The meter will automatically skip the...

Page 21: ...on is restarted with no changes to calibration data CALIBRATION WITH MANUAL TEMPERATURE COMPENSATION Enter the calibration procedure and press LCD to display the temperature on the secondary LCD Unplu...

Page 22: ...r to select code 0 for mV calibration and press CFM to enter CAL will blink on the LCD until the meter confirms a steady reading OFFSET AND SLOPE DIRECT SELECTION Whenever the pH electrode offset and...

Page 23: ...zed at a point near the first calibration point CAL will stop blinking and an intermittent CFM icon will prompt the user to confirm the first calibration If the display stabilizes at a value significa...

Page 24: ...5Vdc 3 0 Vdc 5 Vdc 1 5Vdc 4 1 Vdc 5 Vdc 0 10Vdc 5 0 Vdc 10 Vdc Immerse the temperature probe in the second beaker as near to the Checktemp as possible and repeat the above procedure Calibration proced...

Page 25: ...theEEPROM Date Time Offset in mV for pH 502 only Up to two slopes for pH 502 only Up to three buffers While displaying this data the pH controller remains in con trol mode The procedure below indicate...

Page 26: ...ta is composed of setup data and calibration data As for the setup data the calibration data is assigned default values when an EEPROM reset occurs An un calibrated meter can perform mea surement thou...

Page 27: ...d together the LCD will look like this FAULT CONDITIONS AND SELFTEST PROCEDURES The fault conditions below may be detected by the software EEPROM data error I2C internal bus failure code dead loop EEP...

Page 28: ...the correct position of squares Note A maximum of two keys may be pressed simultaneously to be properly recognized To exit the keyboard test procedure press LCD CAL and SETUP simultaneously EEPROM SE...

Page 29: ...S Temperature has a significant effect on pH The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions For manual temperature calibration please ref...

Page 30: ...for double junction electrodes if necessary Low Slope refer to the cleaning procedure above If the bulb and or junction are dry soak the electrode in HI70300 storage solution for at least one hour Fo...

Page 31: ...absorbs hydrogen in the presence of reducing mediums This phenomenon is rough on the electrode To make correct redox measurements the following condi tions must prevail The surface of the electrode mu...

Page 32: ...3 800 4 740 5 680 6 640 7 580 8 520 9 460 10 400 11 340 12 280 13 220 14 160 Reducing pretreatment immerse the electrode for a few min utes in HI 7091 Oxidizing pretreatment immerse the electrode for...

Page 33: ...HI 3090B 5 BNC connector 5 m 16 5 cable double junction Pt glass body polymer filled RECOMMENDED pH ELECTRODES all electrodes are gel filled and with ce ramic junction unless otherwise indicated HI 1...

Page 34: ...the electrode connector there is also a spade lug connection for the matching pin HI 3210B 5 BNC connector 5 m 16 5 cable double junction Pt plas tic body PVDF junction polymer filled GOLD ORP ELECTRO...

Page 35: ...d Goods Authorization number from the Customer Service department and then send it with ship ping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection H...

Page 36: ...using these products make sure that they are entirely suitable for the environment in which they are used Operation of these instruments in residential areas could cause unaccept able interferences t...

Page 37: ...21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 61...