METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description
Quantity
HI95769A-0
Anionic Surfactants Reagent A 4 drops
HI95769B-0
Anionic Surfactants Reagent B 2 drops
-
Chloroform Reagent
10 mL
DEIONIZED120
Deionized Water
15 mL
REAGENT SETS
HI95769-01
Anionic Surfactants Reagent - 40 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the graduated glass vial with 25 mL of sample.
Note:
For improved accuracy the use of class A
laboratory pipette is recommended.
• Add 2 drops of
HI95769A-0
Anionic Surfactants
Reagent A and 2 drops of
HI95769B-0
Anionic
Surfactants Reagent B.
• Replace the cap and invert to mix, the solution will
turn blue.
• Add 10 mL of Chloroform.
Note:
Chloroform is more dense than water and
will sink to the bottom of the graduated glass vial.
25 mL
×2
×2
10 mL