8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93713-0
Phosphate LR Reagent
1 packet
REAGENT SETS
HI93713-01
Phosphate LR Reagent - 100 tests
HI93713-03
Phosphate LR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Rinse the cuvette by adding a small quantity of sample.
Ensure that the plastic stopper is in place and shake the
cuvette several times. Empty the cuvette.
• Fill the cuvette with 10 mL of sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette and add the content of one
packet of
HI93713-0
Phosphate LR Reagent.
Replace the plastic stopper and the cap. Shake
gently (for about 2 minutes) until the powder
is completely dissolved.
10 mL