7.7.  MARINE NITRATE HR

REQUIRED REAGENTS
Code

Description

Quantity

HI782‑0

Marine Nitrate HR Reagent

1 packet

REAGENT SETS

HI782‑25

Marine Nitrate HR Reagent

25 tests

For other accessories see Accessories section.

MEASUREMENT PROCEDURE

Select the 

Nitrate Marine HR 

method using the procedure described in Method Selection section.

Note

:  If tutorial mode is disabled, follow the measurement procedure below. If the tutorial 

mode is enabled, press 

Measure

 and follow the messages on the screen.

•  Fill the cuvette with 10 mL of unreacted sample (up to the mark).  

Replace the plastic stopper and the cap.

•  Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove.
•  Press 

Zero

. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.

  

  

•  Remove the cuvette.
•  Add the content of one packet of 

HI782‑0 

Marine Nitrate HR Reagent. 

Replace the plastic stopper and the cap. Shake vigorously for 2 minutes.

•  Insert the cuvette into the holder and ensure that the notch on the cap 

is positioned securely in the groove.

•  Press 

Read

. The display will show a 7 minute countdown prior to the measurement. To skip the timer, 

press 

Read

. When the timer ends, the meter will perform the reading. The instrument displays the results 

in 

ppm 

of

 nitrate (NO

3

¯).

  

  

INTERFERENCES

Interference may be caused by:

•  Nitrite

10 mL

Summary of Contents for HI97105

Page 1: ...ive Woonsocket RI 02895 USA www hannainst com HI97105 HI97115 X pH X Alkalinity X Ammonia X Calcium X Magnesium X Nitrate Low Range X Nitrate High Range X Nitrite Ultra Low Range X Phosphate Ultra Low...

Page 2: ...on Selection 17 6 3 Collecting Measuring Samples and Reagents 18 6 4 Cuvette Preparation 18 6 5 Measurement Recommendations 19 6 6 Battery Management 20 7 Method Procedure 21 7 1 Marine pH 21 7 2 Mari...

Page 3: ...Calcium starter kit Reagent A 30 mL bottle 1 pc Reagent B reagent for 25 tests Marine Magnesium starter kit Reagent A 120 mL bottle 1 pc Magnesium Indicator reagent for 25 tests Marine Nitrate High Ra...

Page 4: ...oper disposal of reagent kits and reacted samples 3 SPECIFICATIONS 3 1 PHOTOMETER SPECIFICATIONS Auto logging 200 readings Display 128 x 64 pixel B W LCD with backlight Auto off After 15 minutes of in...

Page 5: ...orimetric EDTA Method using calmagite indicator Reaction between magnesium and the reagents causes a blue to violet tint in the sample Nitrate LR 0 00 to 5 00 ppm as NO3 0 01 ppm 0 25 ppm 2 of reading...

Page 6: ...der in the same position every time With the CAL Check functionality users are able to validate the performance of the instrument at any time Hanna Instruments CAL Check cuvettes are certified against...

Page 7: ...in Romania Woonsocket RI USA 3 x 1 5V AA SN XXXXXXXXXXXX 1 Liquid Crystal Display LCD 2 Keypad 3 ON OFF power button 4 Cuvette holder 5 Indexing mark 6 Battery cover Keypad Description The keypad con...

Page 8: ...tal lattices Photometric chemical analysis is based on specific chemical reactions between a sample and reagent to produce a light absorbing compound If pure absorption occurs the fraction of light ab...

Page 9: ...ions and scratches reducing the need to index the cuvette 5 GENERAL OPERATIONS 5 1 METER VALIDATION CAL CHECK Validation of the photometer involves verifying the concentration of the certified CAL Che...

Page 10: ...100 610 for Alkalinity Ammonia Calcium Magnesium Phosphate ULR then press Next to continue The Please wait message will be displayed during the measurement Note HI97100 ZERO HI97100 525 and HI97100 61...

Page 11: ...ument features a data autolog function to help users keep track of all measurements Every time a measurement is made the data is automatically saved The data log can hold 200 individual measurements N...

Page 12: ...unctional keys to increase or decrease the percentage intensity value Press Accept to confirm or the key to return to the Setup menu without saving the new value Contrast Option 0 to 100 Press Modify...

Page 13: ...n Language Press Modify to change the language Use the functional keys to select the desired language Press Accept to choose one of the languages installed Beeper Option Enable Disable When enabled a...

Page 14: ...isplayed in the top right corner of the screen Meter ID Press Modify and use the functional keys to set desired ID Press Accept to confirm or the key to return to the Setup menu without saving the new...

Page 15: ...y to enter the menu Use the functional keys to select Instruction manual and press Select see details on how to access the online manual Scan the QR code or use link to download the PDF Press the key...

Page 16: ...gured in Setup a 6 digit Bluetooth pin will be generated the first time the instrument and the smart device are paired Once the devices are paired the pin is not required when reconnected Deleting Pai...

Page 17: ...ed method 6 2 READING LOCATION SELECTION The user has the option to select a measurement location from up to 25 TANKS HI97105 and from up to 100 TANKS HI97115 From menu having previously selected requ...

Page 18: ...a vertical position above the cuvette and then push the plunger down into the syringe to deliver desired volume into the cuvette Proper Use of Minipipette 1 Attach the pipette tip 2 Press the button...

Page 19: ...cleaning cloth or a lint free wipe prior to insertion Shaking the cuvette can generate bubbles in the sample causing higher readings To obtain accurate measurements remove such bubbles by swirling or...

Page 20: ...right Always use clean dry cuvettes and syringes tips Rinse with deionized RODI water only never rinse with tank water Dry the cuvettes before use to prevent dilution Clean the syringes and tips befo...

Page 21: ...ollow the messages on the screen Fill the cuvette with 10 mL of unreacted sample up to the mark Replace the plastic stopper and the cap Insert the cuvette into the holder and ensure that the notch on...

Page 22: ...L of unreacted sample up to the mark Replace the plastic stopper and the cap Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press Zero The...

Page 23: ...ent into a clean dry cuvette Caution HI784A 0 is corrosive Dispense liquid slowly and avoid contact with skin and eyes Using scissors open one packet of HI784B 0 Reagent B along the dotted line Push t...

Page 24: ...ow a 15 minute countdown prior to the measurement To skip the timer press Read When the timer ends the meter will perform the reading The instrument displays the ammonia ammonium concentration in ppm...

Page 25: ...arine method using the procedure described in Method Selection section Note If tutorial mode is disabled follow the measurement procedure below If the tutorial mode is enabled press Measure and follow...

Page 26: ...t of HI7582 Marine Calcium Reagent B Replace the plastic stopper and the cap and shake vigorously for 15 seconds or until the powder is completely dissolved Allow air bubbles to dissipate for 15 secon...

Page 27: ...e O rings remain in the tip for a proper seal Use the 5 mL syringe with black printing to measure 4 mL of HI783A 0 reagent Ensure there is no excess reagent on the syringe tip then slowly dispense the...

Page 28: ...meter will perform the reading The instrument displays ppm of Mg 2 Rinse cuvettes caps syringes and tips thoroughly with deionized RODI water and allow to dry completely before storing INTERFERENCES...

Page 29: ...use excessive force Dilution 1 Measure 1 mL of sample using HI740143 1 mL graduated syringe 2 Dispense into mixing vial 3 Add nitrate nitrite free artificial seawater up to the 10 mL mark using HI7401...

Page 30: ...vette has been filled up to the 10 mL mark Replace the plastic stopper and the cap Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press Ze...

Page 31: ...e the plastic stopper and the cap Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press Zero The display will show 0 0 when the meter is ze...

Page 32: ...e up to the mark Replace the plastic stopper and cap Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press Zero The display will show 0 0 w...

Page 33: ...tion section Note If tutorial mode is disabled follow the measurement procedure below If the tutorial mode is enabled press Measure and follow the messages on the screen Fill the cuvette with 10 mL of...

Page 34: ...instrument displays the results in ppm of phosphate PO4 3 Press the key and use the functional keys to select Chemical Form Use the functional keys and press Select to change displayed chemical formul...

Page 35: ...adjust the light level Please check the preparation of the zero cuvette and that the sample does not contain any debris The meter is overheating It cannot operate within published accuracy specificat...

Page 36: ...the oldest Displays before a new log would overwrite the oldest record Press Continue to accept A critical error has occurred Restart the meter If the issue persists please contact Hanna Instruments t...

Page 37: ...6M Deionized water 230 mL HI7101419 Blue carrying case for HI97115C HI97115UC HI731318 Cloth for wiping cuvettes 4 pcs HI731331 Glass cuvette 4 pcs HI731336N Cap for glass cuvette 4 pcs HI731360 Glass...

Page 38: ...city your local household waste disposal service or the place of purchase RECOMMENDATIONS FOR USERS Before using this product make sure it is entirely suitable for your specific application and for th...

Page 39: ...mful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case users are required to correct the interference at thei...

Page 40: ...Zealand ACMA ANNA B1 complies with AS NZS 4268 2012 standard Japan MIC 5 The module complies with the Japanese Technical Regulation Conformity Certification of Specified Radio Equipment ordinance of...

Reviews: