7.2. MARINE ALKALINITY
REQUIRED REAGENTS
Code
Description
Quantity
HI772S
Marine Alkalinity Reagent
1 mL
REAGENT SETS
HI772‑26
Marine Alkalinity Reagent
25 tests
For other accessories see Accessories section.
MEASUREMENT PROCEDURE
Select the
Alkalinity Marine
method using the procedure described in Method Selection section.
Note
: If tutorial mode is disabled, follow the measurement procedure below. If the tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the notch on the cap is
positioned securely in the groove.
• Press
Zero
. The display will show “‑0.0‑” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Use a 1 mL syringe and add 1 mL of
HI772S
Marine Alkalinity Reagent to the sample.
• Replace the plastic stopper and the cap. Invert 5 times to mix.
Note
: Pay attention not to spill reagent otherwise full color development may be inhibited.
• Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove.
• Press
Read
to start reading. The instrument displays the results in
degree
KH (dKH)
.
10 mL
1 mL