background image

1.

GLP/

 key: press to enter 

GLP mode. In calibration mode press to

edit the date and time.

2.

CAL CHECK

 key: press to perform the validation of the meter, or

press and hold for three seconds to enter 

calibration mode.

3.

ZERO/CFM

 key: press to zero the meter prior to measurement, to

confirm edited values or to confirm factory calibration restore.

4.

READ/ /TIMER

 key: In 

measurement mode, press to make a

measurement, or press and hold for three seconds to start a
pre-programmed countdown prior to measurement. In 

GLP mode

press to view the next screen.

5.

ON/OFF

 key: to turn the meter on and off.

6. Liquid Crystal Display (LCD)

7. Cuvette alignment indicator

8. Cuvette holder

DISPLAY ELEMENTS DESCRIPTION:

INSTRUCTION MANUAL

HI 96786

HI 96786

HI 96786

HI 96786

HI 96786

w w w . h a n n a i n s t . c o m

Preliminary examination:

Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each 

HI 96786

 Ion Selective Meter is supplied complete with:

• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual

Note:

Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.

For more details about spare parts and accessories
see “Accessories”.

Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at [email protected].

Functional description:

Light High:

 There is too much light to perform

a measurement. Please check the preparation of
the zero cuvette.

Light Low:

 There is not enough light to perform

a measurement. Please check the preparation of
the zero cuvette.

No Light:

 The instrument cannot adjust the

light level. Please check that the sample does
not contain any debris.

Inverted cuvettes:

 The sample and the zero

cuvette are inverted.

Zero:

 A zero reading was not taken. Follow the

instructions of the measurement procedure for
zeroing the meter.

Under range:

 A blinking 

“0.0”

 indicates that

the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.

Over Range:

 A flashing value of the maximum

concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.

Standard Low:

 The standard reading is less

than expected.

Standard High:

 The standard reading is higher

than expected.

Errors and warnings:

Measurement procedure:

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the LCD
displays dashes, the meter is ready. The
blinking 

“ZERO”

 indicates that the instrument

needs to be zeroed first.

3•

Add to the cuvette 10 mL of sample and
replace the cap.

4•

Place the cuvette into the cuvette holder and
ensure that the notch on the cap is positioned
securely into the groove.

5•

Press 

ZERO/CFM

 and the lamp, cuvette and

detector icons will appear on the display,
depending on the measurement phase.

6•

After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and ready

for measurement. Remove the cuvette.

7•

Add the content of one packet of 

HI 93728-0

reagent.

8•

Replace the cap and immediately shake
vigorously for exactly 10 seconds by moving
the cuvette up and down. Continue to mix
by inverting the cuvette gently and slowly for
50 seconds, while taking care not to induce
air bubbles. A deposit could remain but it
dose not affect the measurement. Time and
way of shaking could sensitively affect the
measurement.

9•

Replace the cuvette into the holder and
ensure that the notch on the cap is positioned
securely into the groove.

10•

Press and hold 

READ/ /TIMER

 for three

seconds and the display will show the
countdown prior to measurement or alternatively
wait for 4 minutes and 30 seconds and press

READ/ /TIMER

. An audible “beep”

indicates the end of countdown period. In all
cases the lamp, cuvette and detector icons
will appear on the display, depending on
the measurement phase.

Measurement 

1. The measuring scheme (lamp, cuvette, detector), appears during

different phases of zero or reading measurement

2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display

ON ZERO READING:

ON SAMPLE READING:

DURING CALIBRATION PROCEDURE:

OTHER ERRORS AND WARNINGS:

Cooling lamp:

 The instrument waits for the

lamp to cool down.

Battery low:

 The battery must be replaced soon.

Dead battery:

 This indicates that the battery

is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.

Cap error:

 Appears when external light enters

in the analysis cell. Assure that the cuvette cap
is present.

Nitrate

ISM

2

9

3

4

5

10

6

7

8

Technical specifications:

Range

0 to 100 mg/L

Resolution

1 mg/L

Accuracy

±5 mg/L ±5% of reading@ 25°C

Typical EMC Dev.

±1 mg/L

Light Source

Tungsten Lamp

Light Detector

Silicon Photocell with narrow band interference
filter @ 525 nm

Method

Adaptation of the cadmium reduction method.
The reaction between nitrate-nitrogen and
the reagent causes an amber tint in the
sample.

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7 ”)

Weight

360 g (12.7 oz.).

Summary of Contents for HI96786

Page 1: ...INSTRUCTION MANUAL HI 96786 HI 96786 HI 96786 HI 96786 HI 96786 www hannainst com Nitrate ISM ...

Page 2: ...for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at tech hannainst com Technical specifications Range 0 to 100 mg L Resolution 1 mg L Accuracy 5 mg L 5 of reading 25 C Typical EMC Dev 1 mg L Light Source Tungsten Lamp Light Detector Silicon Photocell with narrow band interference filter 5...

Page 3: ...nt In GLP mode press to view the next screen 5 ON OFF key to turn the meter on and off 6 Liquid Crystal Display LCD 7 Cuvette alignment indicator 8 Cuvette holder DISPLAY ELEMENTS DESCRIPTION Functional description 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state o...

Page 4: ...er range A blinking 0 0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test Standard Low The stan...

Page 5: ...ng the cuvette gently and slowly for 50 seconds while taking care not to induce air bubbles A deposit could remain but it dose not affect the measurement Time and way of shaking could sensitively affect the measurement 9 Replace the cuvette into the holder and ensure that the notch on the cap is positioned securely into the groove 10 Press and hold READ TIMER for three seconds and the display will...

Page 6: ...e cuvette 7 Place the CAL CHECK Standard HI 96786 11 Cuvette B into the holder and ensure that the notch on the cap is positioned securely into the groove 8 Press CAL CHECK key and the lamp cuvette and detector icons together with CAL CHECK will appear on the display depending on the measurement phase Validation 4 5 7 8 11 At the end of measurement the instrument directly displays concentration in...

Page 7: ...CK Standard HI 96786 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove 5 Press ZERO CFM and the lamp cuvette and detector icons will appear on the display depending on the measurement phase 6 After a few seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration st...

Page 8: ...P to edit the desired day 01 31 If the key is kept pressed the day number is automatically increased Note It is possible to change the editing from day to year and to month by pressing READ TIMER 17 Press ZERO CFM to save the calibration date 18 The instrument displays Stor for one second and the calibration is saved 19 The instrument will return automatically to measurement modeby displaying dash...

Page 9: ...nd accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s battery follow the steps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location and repl...

Page 10: ...vice Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Recommendations for Users Before using these products make sure that they are entirely suitable for your specific application and for the ...

Reviews: