background image

INSTRUCTION MANUAL

HI96785

Honey Color Analyzer

2

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready. 
The blinking 

“CAL”

 indicates that the 

instrument needs to be calibrated first.

3•

  Fill one cuvette with 4 mL of Glycerol, 

up to 5 mm (0.2”) below the rim. This 
is the Glycerol Standard Reference.

4•

  Place the cuvette into the holder paying 

attention to the direction of the light 
indicated by the arrow on the instru-
ment. Then put the light shield cap on 
and ensure that the notch on the cap 
is positioned securely into the groove.

1

3

4

MEASUREMENT PROCEDURE

Measurement

11

5•

 Press 

CAL

 and the lamp, cuvette and 

detector icons will appear on the dis-
play, depending on the measurement 
phase.

6•

  After a few seconds the display will 

show 

“-0.0-”

. The meter is now cali-

brated and ready for measurement.

7•

  Remove the cuvette.

8•

  Add to a second clean cuvette about 

4 mL of honey, up to 5 mm (0.2”) 
below the rim. This is the sample.

9

10

8

5

6

9•

  Place the sample cuvette into the holder 

paying attention to the direction of the 
light indicated by the arrow on the 
instrument. Then put the light shield 
cap on and ensure that the notch on 
the cap is positioned securely into the 
groove.

10• 

Press 

READ

 and the lamp, cuvette and 

detector icons will appear on the  display, 
depending on the measurement phase.

11• 

At the end of measurement, the 
instrument directly displays the honey 
color intensity value expressed in mm 
Pfund, as compared to analytical grade 
Glycerol (fixed at zero mm Pfund).

To save the battery, the instrument shuts down after 10 minutes 
of non-use in measurement mode and after 1 hour of non-use in 
calibration mode. 
If a valid measurement was displayed before 
auto-shut off, the value is displayed when 
the instrument is switched on. The blinking 
“ZERO” means that a new zero has to be 
performed.
One fresh battery lasts for around 750 measurements, depending 
on the light level.
The remaining battery capacity is evaluated at the instrument startup 
and after each measurement.
The instrument displays a battery indicator with three levels as 
follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken 
any more, the instrument shows “dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh 

one.

To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover  

by turning it counterclockwise.

• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.

BATTERY MANAGEMENT

ERRORS AND WARNINGS

On Calibration Reading

INTERFERENCES

Interference may be caused by air bubbles or turbidity in the sample. 
Scratched or dirty cuvettes will also affect readings. Always check 
clearness of cuvettes prior to use.

Light High:

 There is too much light to 

perform a measurement. Please check the 
preparation of the calibration cuvette.

Light Low:

 There is not enough light to 

perform a measurement. Please check the 
preparation of the calibration cuvette.

No Light:

 The instrument cannot adjust the 

light level. Please check that the calibration 
cuvette does not contain any debris.

Inverted cuvettes:

 The sample and the 

calibration cuvette are inverted.

CAL:

 A calibration reading was not taken. 

Follow the instructions of the measurement 
procedure for calibrating the meter.

Under range:

 A blinking 

“0”

 indicates 

that the sample absorbs less light than the 
calibration reference. Check the procedure 
and recalibrate the instrument.

Over Range:

 A flashing value of the 

maximum concentration indicates an over 
range condition (on both wavelenghts). The 
concentration of the sample is beyond the 
programmed range.

A flashing value lower than the maximum 
concentration (e.g. “

120

”) indicates an 

over range condition on one wavelenght. 
The displayed concentration value (e.g. 

120

”) is the reading obtained on the other 

wavelenght.

On Sample Reading:

Cap error:

 Appears when external light 

enters in the analysis cell. Assure that the 
light shield cap is present.

Cooling lamp:

 The instrument waits for the 

lamp to cool down.

Battery low:

 The battery must be replaced 

soon.

Dead battery:

 This indicates that the battery 

is dead and must be replaced. Once this 
indication is displayed, normal operation of 
the instrument will be interrupted. Change 
the battery and restart the meter.

Other Errors and Warnings:

Summary of Contents for HI96785

Page 1: ...INSTRUCTION MANUAL HI96785 Honey Color Analyzer ...

Page 2: ...e information about Hanna Instruments and our products visit www hannainst com For technical support contact your local Hanna Instruments Office or e mail us at tech hannainst com Find your local Hanna Instruments Office at www hannainst com PRELIMINARY EXAMINATION For more details about spare parts and accessories see Accessories Range 0 to 150 mm Pfund Resolution 1 mm Pfund Accuracy 25 C 77 F 2 ...

Page 3: ...e meter on and off 4 Liquid Crystal Display LCD 5 Cuvette holder 1 The measuring scheme lamp cuvette detector appears during different phases of calibration or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass appears when an internal check is in progress 5 Status message 6 Four digit main display ...

Page 4: ...nd commercial classification of monofloral honeys Honey darkens with ageing and other changes in colormayresultfrombeekeeper sinterventionsandfromthedifferent waysofconservation e g useofoldhoneycombs contactwithmetals high temperatures exposition to light etc The primary characteristic for commercial honey classification is color Color classes are expressed in millimeters mm Pfund grades compared...

Page 5: ... 10 6 16 11 62 11 41 41 83 83 11 83 11 11 55 51 27 27 119 35 71 71 110 130 35 119 71 27 83 71 83 Table 2 RECOMMENDATIONS FOR USERS Before using these products make sure that they are entirely suitable for your specific application and for the environment in which they are used Operation of these instruments may cause unacceptable interferences to other electronic equipments this requiring the oper...

Page 6: ... it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase ACCESSORIES WARRANTY Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice All rights are reserve...

Page 7: ... the direction of the light indicated by the arrow on the instru ment Then put the light shield cap on and ensure that the notch on the cap is positioned securely into the groove 1 3 4 MEASUREMENT PROCEDURE Measurement 5 Press CAL and the lamp cuvette and detector icons will appear on the dis play depending on the measurement phase 6 After a few seconds the display will show 0 0 The meter is now c...

Page 8: ...ssed in mm Pfund as compared to analytical grade Glycerol fixed at zero mm Pfund ERRORS AND WARNINGS On Calibration Reading INTERFERENCES Interference may be caused by air bubbles or turbidity in the sample Scratched or dirty cuvettes will also affect readings Always check clearness of cuvettes prior to use Light High There is too much light to perform a measurement Please check the preparation of...

Page 9: ...ond the programmed range A flashing value lower than the maximum concentration e g 120 indicates an over range condition on one wavelenght The displayed concentration value e g 120 is the reading obtained on the other wavelenght On Sample Reading Cap error Appears when external light enters in the analysis cell Assure that the light shield cap is present Cooling lamp The instrument waits for the l...

Page 10: ...t displays a battery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Battery icon blinking if the capacity is under 10 If the battery is empty and accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s...

Reviews: