background image

INSTRUCTION MANUAL

HI96747C

Copper Low Range ISM

Summary of Contents for HI96747C

Page 1: ...INSTRUCTION MANUAL HI96747C Copper Low Range ISM ...

Page 2: ...Hanna Instruments product Please read this instruction manual carefully before using the instrument This manual will provide you with the necessary information for correct use of the instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannainst com Dear Cust...

Page 3: ...CISION AND ACCURACY PRINCIPLE OF OPERATION FUNCTIONAL DESCRIPTION ERRORS AND WARNINGS GENERAL TIPS FOR AN ACCURATE MEASUREMENT STARTUP MEASUREMENT PROCEDURE VALIDATION PROCEDURE CALIBRATION PROCEDURE GLP BATTERY MANAGEMENT BATTERY REPLACEMENT ACCESSORIES 4 5 6 7 10 14 18 24 26 7 9 13 15 20 25 26 5 ...

Page 4: ... Meter is supplied complete with Sample Cuvettes and Caps 2 pcs CAL Check standard cuvettes 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruction Manual and Quick Reference Guide Rigid carrying case Note Save all packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINAT...

Page 5: ... fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrument after 10 minutes of non use in measurement mode or after 1 hour if left in calibration mode The meter uses an exclusive positive locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell It ...

Page 6: ... USEPA approved bicinchoninate method The reaction between Copper and the bicinchoninate reagent causes a purple tint in the sample Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery Type 9V 1 pc Auto Shut off After 10 of non use in measurement mode after 1 hour of non use in calibration mode with last reading reminder Dimensions 192 x 104 x 69 mm 7 6 x 4 1 x 2 7 Weight 320 g 11 3 ...

Page 7: ...ht absorbed depends both on the optical path length through the matter and on the physical chemical characteristics of the substance according to the Lambert Beer Law log I Io ελ c d or A ελ c d Where log I Io Absorbance A Io intensity of incident light beam I intensity of light beam after absorption ελ molar extinction coefficient at wavelength λ c molar concentration of the substance d optical p...

Page 8: ...e has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the calibration zeroing cuvettes are optically identical to provide the same measurement conditions Whenever possible use the same cuvette for both It is necessary that the surface of the cuvette is clean and not scratched This to avoid measurement inter...

Page 9: ...user calibration date or restore factory calibration KEYPAD DESCRIPTION ON OFF to turn the meter on and off ZERO CFM to zero the meter prior to measurement to confirm edited values or to confirm factory calibration restore READ TIMER this is a multi functional key In measurement mode press to make a measurement or press and hold for three seconds to start a pre programmed countdown prior to measur...

Page 10: ... month day and date icons appear when a date is displayed 8 Four digit main display 9 Measuring units 10 Four digit secondary display FUNCTIONAL DESCRIPTION The instrument shows clear messages when erroneous condition appears Messages are also displayed when the obtained values are outside expected range The beeper is playing a beep on errors Light High There is too much light to perform a measure...

Page 11: ...g 0 000 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test ERRORS AND WARNINGS Standard High The...

Page 12: ...lamp The instrument waits for the lamp to cool down d other errors and warnings Battery low The battery must be replaced soon Dead battery This indicates that the battery is dead and must be replaced Once this indication is displayed the meter will lock up Change the battery and restart the meter ERRORS AND WARNINGS ...

Page 13: ... the bottom of this concavity must be at the same level of the 10 mL mark Proper use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same de...

Page 14: ...eading it is important to discard immediately the sample otherwise the glass might become permanently stained All the reaction times reported in this manual are at 25 C 77 F In general the reaction time should be increased for temperatures lower than 20 C 68 F and decreased for temperatures higher than 25 C 77 F In order to maximize accuracy prior to a measurement follow the validation procedure t...

Page 15: ...After the reagents are added the reacted sample is measured When the beeper sounds briefly and the LCD displays dashes the meter is ready The blinking ZERO indicates that the instrument needs to be zeroed first Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap MEASUREMENT PROCEDURE Place the cuvette into the cuvette holder and ensure that the notch on the cap is po...

Page 16: ...d detector icons will appear on the display depending on the measurement phase Add the content of one packet of HI95747 0 Copper Low Range reagent Replace the cap and shake gently for 15 seconds Remove the cuvette MEASUREMENT PROCEDURE Replace the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove ...

Page 17: ...nide Silver For strongly buffered alkaline or acidic samples pH should be adjusted between 6 and 8 before addition of reagent To avoid interferences due to fingerprints oil or dirt it is very important that the cuvette is wiped clean prior to insertion in the cuvette holder Replacement of scratched cuvettes is strongly recommended Alternatively wait for 45 seconds and just press READ TIMER In both...

Page 18: ...hat the notch on the cap is positioned securely into the groove When the beeper sounds briefly and the LCD displays dashes the meter is ready Use the validation procedure to ensure that the instrument is properly calibrated Warning DonotvalidatetheinstrumentwithanystandardsolutionsotherthantheHannaInstruments CAL CheckTM Standards otherwise erroneous results will be obtained For accurate validatio...

Page 19: ...se At the end of the measurement the display will show the validation standard value The reading should be within specifications as reported in the CAL CheckTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are still found out of specifications then recalibrate the instrument ...

Page 20: ...oing Place the CAL CheckTM Standard HI96747 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Note It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions other than the Hanna Instruments CALCheck Standards otherwiseerroneousresul...

Page 21: ...ckTM Standard HI96747 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appear on the display depending on the measurement phase Press ZERO CFM and the lamp cuvette and detector icons will appear on the display depending on the measurement phase CALIBRATION PROCEDURE ...

Page 22: ... 08 2005 appears on the display or 01 01 2005 if the factory calibration was selected before In both cases the year number is blinking ready for date input CALIBRATION PROCEDURE After measurement the instrument will show for three seconds the CAL CheckTM Standard value DATE INPUT Press GLP to edit the desired year 2000 2099 If the key is kept pressed the year number is automatically increased When...

Page 23: ...lly increased When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press ZERO CFM to save the calibration date Note It is possible to change the editing from day to year and to month by pressing READ TIMER The instrument displays Stor for one second and the calibration is saved The instrument will return automatically to measurement...

Page 24: ... display If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns to measurement mode after three seconds LAST CALIBRATION DATE Press READ TIMER to enter in the factory calibration restore screen The instrument asks for confirmation of user calibration delete FACTORY CALIBRATION RESTORE It is possible to delete the calibration...

Page 25: ...hen the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The remaining battery capacity is evaluated at the instrument startup and after each measurement The instrument displays a battery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line...

Page 26: ...lockwise to close To replace the instrument s battery follow the steps BATTERY REPLACEMENT ACCESSORIES Reagent Set HI95747 01 Reagents for 100 tests HI95747 03 Reagents for 300 tests Other Accessories HI96747 11 CAL Check Standard Cuvettes 1 set HI740029P 9V battery 10 pcs HI731318 Cloth for wiping cuvettes 4 pcs HI731331 Glass cuvettes 4 pcs HI731335 Caps for cuvettes HI93703 50 Cuvette cleaning ...

Page 27: ...ts first obtain a Returned Goods Authorization Number from the Customer Service Depart ment and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Recommendations for Users Before using these products make sur...

Reviews: