HI96000 Series

Portable  

Photometers

•  CAL Check

 

·

Allows for performance verification  

  and calibration of the meter using NIST  

  traceable standards.

•  GLP

 

·

 Review of the last calibration date.

•  Auto-shut off

 

·

Automatic shut off after 10 minutes  

  of non-use when the meter is in  

  measurement mode. Prevents wastage  

  of batteries in the event the meter is  

  accidentally left on.

•  Battery status indicator 

 

·

Indicates the amount of battery  

  life left.

•  Built-in timer 

 

·

 Display of time remaining before a  

  measurement is taken. Ensures that  

  all readings are taken at the appropriate  

  reaction intervals for the test being  

 performed.

•  Error messages 

 

·

Messages on display alerting to  

  problems including no cap, high zero,  

  and standard too low.

•  Cooling lamp indicator

 

·

To maintain the desirable wavelength  

  to be used for absorbance, it is necessary  

  to ensure components are not overheated  

  from the heat generated by the tungsten  

  lamp. Each photometer is designed to  

  allow a minimal amount of time for  

  components to cool. The cooling lamp  

  indicator is displayed prior to a reading  

  being taken.

•  Units of measure 

 

·

 Appropriate unit of measure is  

  displayed along with reading. 

Hanna's portable photometers feature an 

advanced optical system; the combination 

of a special tungsten lamp, a narrow band 

interference filter, and silicon photodetector 

ensure accurate photometric readings 

every time. The Hanna exclusive CAL 

Check™ feature utilizes ready-made, NIST 

traceable standards to verify both meter 

validation and calibration. The exclusive 

cuvette locking system ensures that the 

cuvette is inserted into the measurement 

cell in the same position every time to 

maintain a consistent path length.

CAL Check™ Validation*

Two-step validation procedure for proper calibration.

Zero the meter prior to validation... 

 

Place the CAL Check™ Standard A into the cuvette holder and press ZERO/CFM button.   

 

The lamp, cuvette and detector icons will appear on the display followed by “-0.0-”.  

 

The meter is now zeroed and ready for validation.

... and compare accuracy against a known standard. 

 

Place the CAL Check™ standard B into the cuvette holder and press  CAL Check™. The  

 

lamp, cuvette and detector icons together with “CAL Check” will appear on the display.  

 

At the end of the measurement the display will show the validation standard value.

CAL Check™ Calibration*

Calibrate your instrument quickly and easily.

Zero the meter prior to  calibration... 

 

Press and hold CAL Check™ for three seconds to enter calibration mode. Place the  

 

CAL Check™ Standard A into the cuvette holder and press ZERO/CFM. The lamp, cuvette  

 

and detector icons will appear on the display followed by “-0.0-”. The meter is now zeroed  

 

and ready for calibration.

... and calibrate to a known standard. 

 

Place the CAL Check™ Standard B into the cuvette holder. Press READ/TIMER and  

 

the lamp, cuvette and detector icons will appear on the display. After measurement  

 

the instrument will show the CAL Check™ Standard value.

*Each CAL Check™ cuvette is clearly labeled with its respective measurement. Please read the full instruction manual before validation/calibration.

Solutions and Accessories

HI93703-50

Cuvette cleaning solution, 230 mL

HI731335

Cuvette caps (4)

HI731318

Cuvette cleaning cloth (4)

HI740318

Carrying case for HI96 series

HI731331

Measuring cuvettes (4)

"C" versions include  

CAL Check™ standards and  

a rigid carrying case