HI96000 Series
Portable
Photometers
• CAL Check
·
Allows for performance verification
and calibration of the meter using NIST
traceable standards.
• GLP
·
Review of the last calibration date.
• Auto-shut off
·
Automatic shut off after 10 minutes
of non-use when the meter is in
measurement mode. Prevents wastage
of batteries in the event the meter is
accidentally left on.
• Battery status indicator
·
Indicates the amount of battery
life left.
• Built-in timer
·
Display of time remaining before a
measurement is taken. Ensures that
all readings are taken at the appropriate
reaction intervals for the test being
performed.
• Error messages
·
Messages on display alerting to
problems including no cap, high zero,
and standard too low.
• Cooling lamp indicator
·
To maintain the desirable wavelength
to be used for absorbance, it is necessary
to ensure components are not overheated
from the heat generated by the tungsten
lamp. Each photometer is designed to
allow a minimal amount of time for
components to cool. The cooling lamp
indicator is displayed prior to a reading
being taken.
• Units of measure
·
Appropriate unit of measure is
displayed along with reading.
Hanna's portable photometers feature an
advanced optical system; the combination
of a special tungsten lamp, a narrow band
interference filter, and silicon photodetector
ensure accurate photometric readings
every time. The Hanna exclusive CAL
Check™ feature utilizes ready-made, NIST
traceable standards to verify both meter
validation and calibration. The exclusive
cuvette locking system ensures that the
cuvette is inserted into the measurement
cell in the same position every time to
maintain a consistent path length.
CAL Check™ Validation*
Two-step validation procedure for proper calibration.
Zero the meter prior to validation...
Place the CAL Check™ Standard A into the cuvette holder and press ZERO/CFM button.
The lamp, cuvette and detector icons will appear on the display followed by “-0.0-”.
The meter is now zeroed and ready for validation.
... and compare accuracy against a known standard.
Place the CAL Check™ standard B into the cuvette holder and press CAL Check™. The
lamp, cuvette and detector icons together with “CAL Check” will appear on the display.
At the end of the measurement the display will show the validation standard value.
CAL Check™ Calibration*
Calibrate your instrument quickly and easily.
Zero the meter prior to calibration...
Press and hold CAL Check™ for three seconds to enter calibration mode. Place the
CAL Check™ Standard A into the cuvette holder and press ZERO/CFM. The lamp, cuvette
and detector icons will appear on the display followed by “-0.0-”. The meter is now zeroed
and ready for calibration.
... and calibrate to a known standard.
Place the CAL Check™ Standard B into the cuvette holder. Press READ/TIMER and
the lamp, cuvette and detector icons will appear on the display. After measurement
the instrument will show the CAL Check™ Standard value.
*Each CAL Check™ cuvette is clearly labeled with its respective measurement. Please read the full instruction manual before validation/calibration.
Solutions and Accessories
HI93703-50
Cuvette cleaning solution, 230 mL
HI731335
Cuvette caps (4)
HI731318
Cuvette cleaning cloth (4)
HI740318
Carrying case for HI96 series
HI731331
Measuring cuvettes (4)
"C" versions include
CAL Check™ standards and
a rigid carrying case
| www.hannainst.com