Logging
56
6. During active logging, users can track and log notes.
To add a new log note:
• Tap .
• Use on‑screen keypad to enter text.
• Tap
to log entered note. Log Recall view and .CSV log indicate the added log notes.
Autohold with Automatic logging
1. Select
Direct/Autohold
in the Reading Mode (
Reading
tab).
2. Select
Automatic
under Logging Type (
Logging
tab).
3. Place probe in sample.
4. Tap and symbols to initiate the log.
Log points will be stored as soon as is tapped (this includes the changing measurements even when
the meter is trying to find a value to hold at).
5. When Autohold is reached,
Autohold
is displayed and this value will be logged with an “H”in the log notes.
6. Tap
to stop the log and tap
DO
to restore direct measurements.
Note:
Logged data is recorded up to the Autohold changing and every point after is the held value until the
log was stopped (i.e. 1,2,3,4,5,5,5,5,5). The Autohold value is identified with an “H” in the log record.
12.2. MANUAL LOGGING
1. From Measurement screen, tap (Measurement Menu).
2. Tap
Logging
tab and select
Manual
logging type.
3. See option to create file name:
• Tap
Create
.
• Use on‑screen keypad to enter file name.
• Tap
to save new name.
4. See option to define sample ID (name and prefix).
5. Tap (
Home
key) to reenter Measurement screen.
6. From Measurement screen, tap
M
(manual logging symbol) to log data.
Data is logged every time symbol is tapped.
7. From Measurement screen, tap symbol to start/stop logging.