background image

Range (*)

-2.00 to 16.00 pH

-5.0 to 105.0 °C / 23.0 to 221.0 °F

Resolution

0.01 pH

0.1 °C  /  0.1 °F

Accuracy

±0.02 pH

(@25 °C/77 °F)

±0.5 °C up to 60 °C; ±1.0 °C outside

                            ±1.0 °F up to 140 °F; ±2.0 °F outside

Temperature Compensation                 

Automatic

pH Calibration

Automatic, 1 or 2 point

choose between 2 sets of buffers

(standard: 4.01/7.01/10.01 or NIST: 4.01/6.86/9.18)

Probe (included)

FC 215D

preamplified pH/temperature

probe with DIN connector and 1 m (3.3’) cable

Battery Type

3 x 1.5V AAA

Battery Life

Approximately 1200 hours of continuous use

Auto-off

After 8 minutes of non-use

Environment

0 to 50 °C (32 to 122 °F);

RH max. 100%

Dimensions

152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)

Weight

205 g (7.2 oz.)

(*)

 the temperature range is limited from 0 to 50 °C (32 to

122 °F) when using 

FC 215D

 sensor.

To clean the meter, use water only.

HI 99191

 is guaranteed for two years against defects in

workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodes
are guaranteed for a period of six months. This warranty is
limited to repair or replacement free of charge.     Damage due to
accidents, misuse, tampering or lack of prescribed maintenance
is not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the
model number, date of purchase, serial number and the
nature of the problem. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization number from the Technical Service
department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly
packaged for complete protection.

Remove the instrument from the packing material and examine
it carefully to make sure that no damage has occurred during
shipment. If noticeable damage is evident, notify your dealer.

Note:

Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing together with the
supplied accessories.

HI 99191

     is a portable, microprocessor-based pH and temperature

meter specifically designed for Drinking Water Analysis.
Main features include: waterproof and compact casing; large
dual-line display; low battery detection; automatic pH calibration
at one or two points within two memorized buffer sets
(standard or NIST); selectable temperature unit (°C or °F).
The pH electrode has a built-in temperature sensor for simultaneous
temperature compensated pH and temperature readings, and
also contains an integral pH sensor preamplifier to provide
measurements impervious to noise and electrical interferences.

FC 215D

 pH electrode with a built-in temperature sensor,

features a rugged, easy to clean, PVDF body and a special
sensitive glass membrane which allows a quick response time in
low conductivity samples as drinking water is.
Each meter is supplied with:

FC 215D

 pH electrode with internal temperature sensor, for

Drinking Water Analysis, with DIN connector and 1 m (3.3’)
cable

• pH 4.01 & pH 7.01 buffer sachets

HI 700661

 electrode cleaning solution (2x20 mL)

• 3 x 1.5V AAA batteries
• rugged carrying case
• instruction manual

Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance
notice.

FC  215D

pH electrode with internal temperature sensor, for
Drinking Water Analysis, with DIN connector and 1 m
(3.3’) cable

HI  7004L

pH 4.01 buffer solution, 500 mL

HI  7006L

pH 6.86 buffer solution, 500 mL

HI  7007L

pH 7.01 buffer solution, 500 mL

HI  7009L

pH 9.18 buffer solution, 500 mL

HI  7010L

pH 10.01 buffer solution, 500 mL

HI  70300L

pH electrode Storage solution, 500 mL

HI  7061L

pH electrode Cleaning solution, 500 mL

HI  70630L

pH electrode Acid Cleaning solution 500 mL

HI  70631L

pH electrode Alkaline Cleaning solution, 500 mL

HI  70632L

pH electrode Cleaning and Disinfection solution, 500 mL

HI  7071

Refilling Solution

1. Automatic Temperature Compensation indicator
2. Battery percentage (visible at power up)
3. Secondary display
4. Low battery indicator
5. Stability indicator
6. pH calibration indicator
7. Selectable temperature unit
8. Primary display
9. Measuring unit for primary display

IST99191    06/13

WARRANTY

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

ACCESSORIES

LCD DESCRIPTION

SPECIFICATIONS

Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using this
instrument. This manual will provide you with the necessary
information for correct use of these instruments, as well as a
precise idea of their versatility. If you need additional technical
information, do not hesitate to e-mail us at
[email protected] or view our worldwide contact list at
www.hannainst.com.

All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner.

Instruction Manual

HI 99191

Portable Waterproof pH / Temperature

Meter for Drinking Water Analysis

Reviews: