Setup mode allows viewing and configuring the following parameters:
• Beep Status (bEEP)
• Auto power off timer (AOFF)
• Temperature Unit
• Calibration Stability Criteria
To enter SETUP mode, press
Setup
while
the instrument is in measurement mode.
Select the desired setup parameter us-
ing the
STD
key.
Press
CAL
if you want to change the
item value. The selected item starts
blinking for about 8 seconds.
Press the
Select
key to change the
displayed value.
Wait until item stops blinking to complete selection or press
CAL
to
escape.
Press the
STD
key to select the next parameter.
Press
Setup
to exit SETUP menu at any time.
The following table lists the SETUP parameters, their valid values
range and the factory settings (default).
Notes:
• Press and hold
Setup
then
Select
before
“CFM”
tag
appears, to toggle between standard and temperature
display on secondary LCD line reading during calibration.
• If the value measured by the instrument is not close to the
selected standard, “
WRONG
” “ ” and “
WRONG
” “ ” tags
will blink alternately. In this case check if the correct standard
has been selected or used, or regenerate the electrode and the
reference by following the Cleaning Procedure (see page 16). If
necessary, change the standard, the electrode, or the reference.
Press
CFM,
then
CAL
before
“CFM”
tag appears in order to
clear previous calibration and set the default values.
The
“CLR ALL”
message is displayed on both LCD lines and
the instrument returns to measurement mode.