Battery Replacement

Battery life percentage indicator is displayed at power on. If the 

battery level drops below 5%, the   symbol lights up indicating 

that the batteries should be replaced soon.

If the battery level is not adequate to guarantee an accurate reading, 

the tester automatically turns off. 

To replace the batteries, follow the next steps:
1.  Turn OFF the tester.
2.  Remove the four screws on the top of the tester to open the battery 

compartment (figure 1).

3.  Remove the old batteries.
4.  Insert four new 1.5V AA batteries in the battery compartment 

while paying attention to the correct polarity (figure 2).

5.  Close the battery compartment using the four screws.

fig. 1

fig. 2

Note: 

Only use the battery type specified in the manual. Old 

batteries should be disposed in accordance with local regulations.

Accessories

Electrode

HI73127

Replaceable pH electrode

HI73128

Electrode removal tool

pH Buffer Solution

HI70004P

pH 4.01 buffer solution, 20 mL sachet (25 pcs.)

HI70006P

pH 6.86 buffer solution, 20 mL sachet (25 pcs.)

HI70007P

pH 7.01 buffer solution, 20 mL sachet (25 pcs.)

HI70009P

pH 9.18 buffer solution, 20 mL sachet (25 pcs.)

HI70010P

pH 10.01 buffer solution, 20 mL sachet (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachet 

(10 pcs., 5 each)

HI770710P

pH 10.01 & 7.01 buffer solution, 20 mL sachet 

(10 pcs., 5 each)

ORP Test Solution

HI7021M

ORP test solution (240 mV), 230 mL

HI7022M

ORP test solution (470 mV), 230 mL

Pretreatment Solution

HI7091L

Pretreatment reducing solution

HI7092M

Pretreatment oxidizing solution, 230 mL

Electrode Cleaning Solution

HI7061M

General purpose cleaning solution, 230 mL

Electrode Storage Solution

HI70300M

Electrode storage solution, 230 mL

Other Accessories

HI740026P

Replacement 1.5 V Batteries (12 pcs.)

pH Measurement

Press SET/HOLD to select pH mode. Place 
the electrode in the solution to be tested 
while stirring gently. The measurement 
should be taken after the stability tag   
has disappeared. The pH value is displayed on the first LCD line and 
the temperature on the second LCD line.

Note: 

Before taking any pH measurement make sure the tester 

has been calibrated (“CAL“ tag is displayed).

ORP Measurement

Use SET/HOLD to select ORP mode. Place 
the electrode in the solution to be tested. 
The measurement should be taken after the 
stability tag   has disappeared.
The ORP (mV) value is displayed on the first LCD line and the 
temperature on the second LCD line.

Maintenance

Please read the information below to ensure the highest possible 
accuracy.
• Fresh buffer solution should be used for each calibration.
• For improved accuracy a two-point calibration is recommended.
• If measurements are taken successively, rinse the probe thoroughly  

in distilled or deionized water to eliminate cross-contamination.

• Clean the electrode monthly using pH and ORP cleaning solution. 

Keep the electrode in 

HI700601

 - pH and ORP electrode cleaning 

Solution for 30 minutes and rinse the probe thoroughly with water.

• When not in use, add a few drops of 

HI70300

 Storage solution to 

the protective cap. Never store the probe in distilled or deionized 
water.

Electrode Replacement

Use supplied removal tool (

HI73128

) to replace the pH electrode.

Insert the tool into the 
probe cavity.

Rotate the tool 
counterclockwise.

Pull the electrode out by 
using the other side of 
the tool.

Insert a new pH electrode following the above instructions in reverse 
order.

Certification

All Hanna Instruments conform to the 

CE European 

Directives.
Disposal of Electrical & Electronic Equipment.

 The 

product should not be treated as household waste. 
Instead hand it over to the appropriate collection point 
for the recycling of electrical and electronic equipment 
which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do not 

dispose of them with other household waste. Hand them over to the 
appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential 
negative consequences for the environment and human health. For 
more information, contact your city, your local household waste 
disposal service, the place of purchase or go to www.hannainst.com.

Recommendations for Users

Before using this tester, make sure that it is entirely suitable for 
your specific application and for the environment in which it is used. 
The glass bulb at the end of the electrode is sensitive to electrostatic 
discharges. Avoid touching the glass bulb and the ORP electrode at all 
times. Any variation introduced by the user to the supplied equipment 
may degrade the tester’s performance. For yours and the tester’s 
safety do not use or store the tester in hazardous environments.

Warranty

HI98121

 is warranted for a period of one year against defects in 

workmanship and materials when used for its intended purpose and 
maintained according to instructions. The electrode is warranted for a 
period of six months. This warranty is limited to repair or replacement 
free of charge. Damage due to accidents, misuse, tampering or lack of 
prescribed maintenance is not covered. If service is required, contact 
your local Hanna Instruments Office. If under warranty, report the 
model number, date of purchase, serial number and the nature of 
the problem. If the repair is not covered by the warranty, you will be 
notified of the charges incurred. 
If the instrument is to be returned to Hanna Instruments Office, 
first obtain a Returned Goods Authorization (RGA) number from the 
Technical Service department and then send it with shipping costs 
prepaid. When shipping any instrument, make sure it is properly 
packaged for complete protection.

Hanna Instruments reserves the right to modify the design, construction, 

or appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the 

written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, 

Rhode Island, 02895, USA.

IST98121 09/19

HI98121

Waterproof pH, ORP 

& Temperature Tester

INSTRUCTION MANUAL

Summary of Contents for HI 98121

Page 1: ...HI98121 Waterproof pH ORP Temperature Tester INSTRUCTION MANUAL ...

Page 2: ...on manual Note Save all packing material until you are sure that the tester works correctly Any damaged or defective item must be returned in its original packing material with the supplied accessories General Description Intended Use HI98121isacompactpH ORPandtemperaturetester Itfeaturesatwo buttonsoperationsystemandiseasytouse Thecompactandwaterproof casing is designed to float if accidentally d...

Page 3: ...of standard buffers pH 4 01 7 01 10 01 or pH 4 01 6 86 9 18 ORP factory calibrated Electrode HI73127 pH electrode included Environment 5 to 50 C 23 to 122 F RH 100 Battery type 1 5V 4 pcs Battery life Approx 250 hours Auto off After 8 minutes of non use Dimensions 171 x 41 x 26 mm 6 7 x 1 6 x 1 0 Weight 83 g 2 9 oz Functional Description Liquid Crystal Display SET HOLD button Battery compartment M...

Page 4: ...hold SET HOLD until HOLD is displayed pH or ORP readings will be frozen on the LCD Press either buttons to resume measurement mode Setup Setup mode allows the selection of temperature unit and pH buffer set To enter the setup mode select pH mode and press MODE until TEMP and the current temperature unit are displayed e g TEMP C Then Press SET HOLD to select temperature unit After the temperature u...

Page 5: ...calibration using pH 7 01 or 6 86 buffer solution press MODE after calibration point has been accepted The tester will display pH 7 01 or pH 6 86 and OK 1 and then return to measurement mode For a two point calibration using pH 7 01 or pH 6 86 buffer solution after the first calibration point has been accepted pH 4 01 USE will be displayed for 12 seconds unless a valid buffer is recognized If vali...

Page 6: ...ure the highest possible accuracy Fresh buffer solution should be used for each calibration For improved accuracy a two point calibration is recommended If measurements are taken successively rinse the probe thoroughly in distilled or deionized water to eliminate cross contamination Clean the electrode monthly using pH and ORP cleaning solution Keep the electrode in HI700601 pH and ORP electrode c...

Page 7: ...in accordance with local regulations Accessories Electrode HI73127 Replaceable pH electrode HI73128 Electrode removal tool pH Buffer Solution HI70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI70006P pH 6 86 buffer solution 20 mL sachet 25 pcs HI70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI70009P pH 9 18 buffer solution 20 mL sachet 25 pcs HI70010P pH 10 01 buffer solution 20 mL sachet...

Page 8: ... performance For yours and the tester s safety do not use or store the tester in hazardous environments Warranty HI98121 is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions The electrode is warranted for a period of six months This warranty is limited to repair or replacement free of charge D...

Reviews: