background image

1.  

GLP/  

   key: press to enter 

GLP mode

. In 

calibration mode

 press to edit

the date and time.

2.

CAL CHECK 

key: press to perform the validation of the meter, or press

and hold for three seconds to enter 

calibration mode

.

3.

ZERO/CFM

 key: press to zero the meter prior to measurement, to

confirm edited values or to confirm factory calibration restore.

4.

READ  /TIMER

 key: In 

measurement mode

, press to make a

measurement, or press and hold for three seconds to start a pre-
programmed countdown. In 

GLP mode

 press to view the next screen.

5.

ON/OFF

 key: to turn the meter on and off.

6. Liquid Cristal Display (LCD)
7. Cuvet alignment indicator
8. Cuvet holder

3

DISPLAY ELEMENTS DESCRIPTION:

INSTRUCTION MANUAL

HI 96770

HI 96770

HI 96770

HI 96770

HI 96770

Silica High Range ISM

This instrument is in

compliance with the

Directives

www.hannainst.com

Preliminary examination:

Please examine this product carefully. Make sure that the instrument is not

damaged. If any damage occurred during shipment, please notify your

Dealer.

Each 

HI 96770

 Ion Selective Meter is supplied complete with:

• Two Sample Cuvets and Caps
• 9V Battery
• Instruction Manual

Note:

save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.

For more details about spare parts and
accessories see “Accessories”

Dear Customer,

Thank you for choosing a Hanna product. This manual will provide you

with the necessary information for the correct use of the instrument. Please

read it carefully before using the meter. If you need additional technical

information, do not hesitate to e-mail us at [email protected].
This instrument is in compliance with   directives.

Technical specifications:

Range

0 to 200 mg/L (as SiO

2

)

Resolution

1 mg/L for 

measurement

Precision

±5 mg/L @ 100 mg/L

Typical EMC Dev.

±1 mg/L

Light Source

Light Emitting Diode

Light Detector

Silicon Photocell with narrow band
interference filter @ 466 nm

Method

Adaptation of the USEPA METHOD 370.1
for drinking, surface and saline waters,
domestic and industrial wastes and Standard
Method 4500-SiO

2

 C.

Environment

0 to 50°C (32 to 122°F);
max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode

;

after 1 hour of non-use in 

calibration mode

;

with last reading reminder.

Dimensions

192 x 102 x 67 mm (7.6 x 4 x 2.6”)

Weight

290 g (10 oz.).

Functional description:

Light High:

 There is too much light to perform a

measurement. Please check the preparation of the
zero cuvet.

Light Low:

 There is not enough light to perform a

measurement. Please check the preparation of the
zero cuvet.

No Light:

The instrument cannot adjust the light

level. Please check that the samples does not contain
any debris.

Inverted cuvets:

 The sample and the zero cuvet

are inverted.

Zero:

 A zero reading was not taken. Follow the

instructions of the measurement procedure for zero-
ing the meter.

Under range:

 A blinking 

“0”

 indicates that the

silica content in the sample is less than 0 mg/L.

Over Range:

 A flashing value of the maximum

concentration indicates an over range condition. The
concentration of the sample is beyond the
programmed range: dilute the sample and re-run
the test.

Standard Low:

 The standard reading is out of

range.

Standard High:

 The standard reading is out of

range.

Errors and warnings:

Measurement procedure:

1•

Turn the meter on by pressing

ON/OFF

.

2•

When the beeper sounds and the LCD
displays dashes, the meter is ready.
The blinking 

“ZERO”

 indicates that

the instrument needs to be zeroed first.

3•

Fill the cuvet with 10 mL of unreacted

sample, up to the mark, and replace the

cap.

4•

Place the cuvet into the cuvet holder

and ensure that the notch on the cap

is positioned securely into the groove.

5•

Press 

ZERO/CFM

 and the instrument

status indicator will appear on the

display, depending on the measurement

phase.

6•

After a few seconds the display will

show 

“-0.0-”

. The meter is now

zeroed and ready for measurement.

7•

Remove the cuvet.

8•

Add the content of one packet of
HI 96770A-0 Silica HR reagent A
and shake until complete dissolution.

9•

Remove the cap and add the content
of one packet of HI 96770B-0 Silica
HR reagent B. Replace the cap and
shake  vigorously until complete

vigorously until complete

vigorously until complete

vigorously until complete

vigorously until complete

dissolution

dissolution

dissolution

dissolution

dissolution.

10•

Press and hold READ 

READ 

READ 

READ 

READ  /TIMER

/TIMER

/TIMER

/TIMER

/TIMER for

three seconds. The display will show
the countdown. The beeper indicates
the end of countdown period.
Alternatively, wait for 10 minutes.
Remove the cap.

11•

Add the content of one packet of
HI 96770C-0 Silica HR reagent C
and shake vigorously until complete
dissolution.

12•

Replace the cuvet into the cuvet holder
and ensure that the notch on the cap
is positioned securely into the groove.

Measurement 

1) Instrument status indicator appears during different phases of zero

or reading measurement

2) Icons displayed in this area are warning or action icons, signaling

next operation

3) The battery icon shows the charging level of the battery

4) The hourglass appears when the instrument is performing an internal

check up

5) CAL: appears when the instrument operates in the 

calibration mode

6) The month, day and date icons appear when a date is displayed

7) The chronometer appears when the reaction timer is running

8) Four digit main display

9) Measuring units

10) Four digit secondary display

ON ZERO READING:

ON SAMPLE READING:

DURING CALIBRATION PROCEDURE:

OTHER ERRORS AND WARNINGS:

Dead battery:

 This indicates that the battery is

dead and must be replaced. Once this indication
is displayed, the meter will lock up. Change the
battery and restart the meter.

Cap error:

 Appears when external light enters in

the analysis cell. Assure that the cuvet cap is present.

2

4, 12

5-6

10

8, 9

Cooling lamp:

 The instrument waits for the lamp

to cool down.

Battery low:

 The battery voltage is getting low

and the battery needs to be replaced.

11