background image

1.

RANGE/GLP/

 key: press to change the range or press and hold

for three seconds to enter 

GLP mode. In calibration mode press to edit

the date and time.

2.

CAL CHECK

 key: press to perform the validation of the meter, or

press and hold for three seconds to enter 

calibration mode.

3.

ZERO/CFM

 key: press to zero the meter prior to measurement, to

confirm edited values or to confirm factory calibration restore. Press
and hold for three seconds to start a preprogramed countdown prior
to measurement.

4.

READ/ /UNIT

 key: In 

measurement mode, press to make a

measurement, or press and hold for three seconds to change the
measurement unit. In 

GLP mode press to view the next screen.

5.

ON/OFF

 key: to turn the meter on and off.

6. Liquid Crystal Display (LCD)

7. Cuvette alignment indicator

8. Cuvette holder

DISPLAY ELEMENTS DESCRIPTION:

INSTRUCTION MANUAL

HI 96735

HI 96735

HI 96735

HI 96735

HI 96735

w w w . h a n n a i n s t . c o m

Preliminary examination:

Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each 

HI 96735

 Ion Selective Meter is supplied complete with:

• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual

Note:

Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.

Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the meter.
This manual will provide you with the necessary information for correct use
of the instrument. If you need additional technical information, do not
hesitate to e-mail us at [email protected] or view our worldwide
contact list at www.hannainst.com.

Functional description:

Light High:

 There is too much light to perform

a measurement. Please check the preparation of
the zero cuvette.

Light Low:

 There is not enough light to perform

a measurement. Please check the preparation of
the zero cuvette.

No Light:

 The instrument cannot adjust the

light level. Please check that the sample does
not contain any debris.

Inverted Cuvettes:

 The sample and the zero

cuvette are inverted.

Zero:

 A zero reading was not taken. Follow the

instructions of the measurement procedure for
zeroing the meter.

Under Range:

 A blinking 

“200”

 indicates that

the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.

Over Range:

 A flashing value of the maximum

concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.

Standard Low:

 The standard reading is less

than expected.

Standard High:

 The standard reading is higher

than expected.

Errors and warnings:

Measurement procedure:

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the LCD
displays dashes and “P1” (Low range), “P2”
(Medium range) or “P3” (High range), the
meter is ready. The code that appears on
the secondary display is the one of the
last selected range. If necessary, press

RANGE/GLP/

 to change range. The

blinking 

“ZERO”

 indicates that the

instrument needs to be zeroed first.

3•

Add 0.5 mL of unreacted sample to the
cuvette. Add 0.5 mL of 

HI 93735IND-0

indicator reagent. With the plastic dropper
fill the cuvette up to the 10 mL mark
adding 

HI 93735A-XR

 (X=L, M, H)

reagent appropriate to the selected range.

4•

Add two drops of 

HI 93735B-0

 buffer

reagent. Replace the cap and invert 5 times
to mix.

5•

Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.

6•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

7•

After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and ready

for measurement. Remove the cuvette.

8•

Add the content of 1 packet of 

HI 93735C-0

fixing reagent. Replace the cap and shake
gently to mix 20 seconds.

9•

 Replace the cuvette into the holder and

ensure that the notch on the cap is positioned
securely into the groove.

10•

Press 

READ/ /UNIT

. In all cases the lamp,

cuvette and detector icons will appear on the
display, depending on the measurement phase.

Measurement 

1. The measuring scheme (lamp, cuvette, detector), appears during

different phases of zero or reading measurement

2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display

ON ZERO READING:

ON SAMPLE READING:

DURING CALIBRATION PROCEDURE:

OTHER ERRORS AND WARNINGS:

Battery Low:

 The battery must be replaced

soon.

Dead Battery:

 This indicates that the battery

is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.

Cap Error:

 Appears when external light enters

in the analysis cell. Assure that the cuvette cap
is present.

Cooling lamp:

 The instrument waits for the

lamp to cool down.

Hardness ISM

3

6

9

10

10 mL

4

5

8

Technical specifications:

Range

LR        0 to 250 mg/L
MR       200 to 500 mg/L
HR       400 to 750 mg/L

Resolution

1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 750 mg/L

Accuracy

LR   ±5mg/L ±4% of reading@ 25°C
MR  ±7 mg/L ±3% of reading@ 25°C
HR  ±10 mg/L ±2% of reading@ 25°C

Light Source

Light Emitting Diode

Light Detector

Silicon Photocell with narow band interference
filter @ 466 nm

Method

Adaptation of the EPA recommended method
130.1. The reaction between calcium,
magnezium and the reagents causes a
red-violet tint in the sample.

Environment

0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)

Weight

320 g (11.3 oz.).