Hanna Instruments HI 95734C Instruction Manual Download Page 12

12

• In order to measure exactly 5 mL of reagent with the 5 mL syringe:

(a) push the plunger completely into the syringe and insert the tip into the reagent bottle.
(b) pull the plunger up until the lower edge of the seal is exactly on the 5 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are

hanging on the tip of the syringe, if so eliminate them. Then, keeping the syringe in
vertical position above the cuvet, push the plunger completely down into the syringe. Now
the exact amount of 5 mL has been added  to the cuvet.

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• For a correct filling of the cuvet: the liquid in the

cuvet forms a convexity on the top; the bottom of

this convexity must be at the same level of the 10
mL mark.

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

(a)

(b)

(c)

13

• Proper use of the dropper:

(a) to get good reproducible results, tap the dropper on the table for several times and wipe  the

outside of the dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour  out the content of the packet.

• Shaking the cuvet can generate bubbles in the sample, causing higher readings. To obtain accurate

measurements, remove such bubbles by swirling or by gently tapping the vial.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass might

become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule of

thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to be

sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

(a)

(b)

(c)

(a)

(b)

Summary of Contents for HI 95734C

Page 1: ...Instruction Manual HI 95734C Free Total Chlorine High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 24 MAN95734R3 07 06 w w w h a n n a i n s t c o m...

Page 2: ...RE 18 ACCESSORIES 20 BATTERY REPLACEMENT 21 DECLARATION OF CONFORMITY 21 WARRANTY 22 HANNA LITERATURE 22 USER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you...

Page 3: ...ooks and leaflets contact your dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com HI 95734 is warranted for t...

Page 4: ...glass to obtain best results The HI 95734 meter measures the free and total chlorine Cl2 content in water samples in the 0 00 to 10 00 mg L ppm range The method is an adaption of the USEPA METHOD 330...

Page 5: ...result to the true value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 3 00 mg L...

Page 6: ...of Hanna s HI 95 series colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical...

Page 7: ...ot to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination...

Page 8: ...holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed a...

Page 9: ...3734C 0 reagent to the cuvet For both free and total chlorine add the content of one packet of HI 93701 0 Free Chlorine reagent Replace the cap and shake gently for 20 seconds or 2 minutes in case of...

Page 10: ...ervice Center CALIBRATION MODE MESSAGES The date of the last calibration appears on the LCD each time the calibration mode is entered If calibration is performed for the first time F CAL appears This...

Page 11: ...surement A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flashing...

Page 12: ...lint free cloth prior to insertion GENERAL TIPS FOR AN ACCURATE MEASUREMENT a b c 13 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wip...

Reviews: