ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-to-
noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check
that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.

Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check if
the right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in
the 

measurement procedure

 for zeroing the meter.

Under range. A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make
sure you use the same cuvet for reference (zero) and measurement.

A flashing value of the maximum concentration indicates an over
range condition. The concentration of the sample is beyond the

programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates
a  low signal-to-noise ratio condition. In this case accuracy of the
result is not guaranteed. Repeat the 

measurement procedure

.

14

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “734f” or “734t” it is

ready. The display code that appears is the one
of the last selected parameter.

• keep ZERO/SET pressed to change parameter.

The parameter code will change every three
seconds, until the key is released. Select the
program for free chlorine “734f” or total chlorine
“734t”.

• After the desired parameter code appears on

the display, perform either MEASUREMENT,
VALIDATION, or CALIBRATION PROCEDURE
following the instructions reported in the related
chapter.

Code

Parameter

734f

Free Chlorine

734t

Total Chlorine

or

Important Note

: Free and Total Chlorine have to be measured separately with fresh unreacted

samples if both values are requested.

OPERATIONAL GUIDE

Summary of Contents for HI 95734C

Page 1: ...Instruction Manual HI 95734C Free Total Chlorine High Range ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 24 MAN95734R3 07 06 w w w h a n n a i n s t c o m...

Page 2: ...RE 18 ACCESSORIES 20 BATTERY REPLACEMENT 21 DECLARATION OF CONFORMITY 21 WARRANTY 22 HANNA LITERATURE 22 USER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you...

Page 3: ...ooks and leaflets contact your dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com HI 95734 is warranted for t...

Page 4: ...glass to obtain best results The HI 95734 meter measures the free and total chlorine Cl2 content in water samples in the 0 00 to 10 00 mg L ppm range The method is an adaption of the USEPA METHOD 330...

Page 5: ...result to the true value Although good precision suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 3 00 mg L...

Page 6: ...of Hanna s HI 95 series colorimeters is based on special subminiature tungsten lamps and narrow band interference filters to guarantee both high performance and reliable results Block diagram optical...

Page 7: ...ot to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination...

Page 8: ...holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed a...

Page 9: ...3734C 0 reagent to the cuvet For both free and total chlorine add the content of one packet of HI 93701 0 Free Chlorine reagent Replace the cap and shake gently for 20 seconds or 2 minutes in case of...

Page 10: ...ervice Center CALIBRATION MODE MESSAGES The date of the last calibration appears on the LCD each time the calibration mode is entered If calibration is performed for the first time F CAL appears This...

Page 11: ...surement A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flashing...

Page 12: ...lint free cloth prior to insertion GENERAL TIPS FOR AN ACCURATE MEASUREMENT a b c 13 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wip...

Reviews: