
42
The printout defines functions A, B and C:
A
is a start time,
B
is the time interval,
C
is the interrupt value of
pH/mV.
Hours and minutes must be
separated by a decimal
point. The minimum time in-
terval is 1 minute.
E.g. Starting time 10:30, a printout interval of 1h and 30 minutes, pH
interruption value 3.50.
If you need to interrupt the program before the endpoint is reached,
press the 0 key plus the PGM key or enter into any other program.
PROGRAM #6 (HI9318)
To enter the program press the 6
key plus the PGM key.
This program allows you to
set upper and lower limits
for pH/mV values.
A beeper will sound when-
ever the measured value
exceeds the limit.
PROGRAM 5
A enter start
time
B enter time in-
terval
C enter end
pH/mV value
A
C
B
1
1
0
3
.
.
.
3
5
3
0
0
0
1)
2)
3)
6
PGM
PROGRAM 6
A set lower limit
B set upper limit
C clear alarm