1
Instruction Manual
HI 8064
Portable Thermohygrometer
Page 1: ...1 InstructionManual HI 8064 Portable Thermohygrometer...
Page 2: ...ted to allow air circulation and thus guarantee a fast response time To measure humidity electronically a thin film polymer ca pacitance is used TheprobehandleismadeoftoughABSmaterial towithstandthewe...
Page 3: ...ainer Pour approximately 26 cc of distilled water into the other chamber marked RH 75 4 andaddthecontentsoftheenvelopecontain ingNaCl whileshakingthecontainertoavoid any lump formation Note When not i...
Page 4: ...tightlysealthecham ber containing the LiCl solution Remove the plug from the second chamber and insert the probe Waitforthereadingtostabilizeandthennote the displayed value By using the Calibration Ta...
Page 5: ...9 8 RELATIVE HUMIDITY CALIBRATION TABLE...
Page 6: ...according to the instructions The probes are warranted for a period of six months Damages due to accident misuse tamper ing or lack of prescribed maintenance are not covered This warranty is limited...