HI 739 Fluoride High Range

Quick Reference Guide

w w w . h a n n a i n s t . c o m

TECHNICAL SPECIFICATIONS

Range

:

0.0 to 20.0 ppm

Resolution

:

0.1 ppm

Accuracy

:

±0.5 ppm ±5 % of reading @ 25 °C / 77 °F

Typical EMC Dev.

:

±0.1 ppm

Light Source

:

Light Emitting Diode @ 575 nm

Light Detector

:

Silicon Photocell

Method

:

Adaptation of the 

Standard Methods for the Examination of

Water and Wastewater, 18

th

 edition

, SPADNS method.

Environmental

:

0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing

Battery Type

:

1 x 1.5V AAA

Auto-Shut off

:

After 10 minutes of non-use

Dimensions

:

81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5 ”)

Weight

:

64 g (2.25 oz.)

TIPS FOR AN ACCURATE MEASUREMENT

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and

completely free of fingerprints, oil and dirt. Wipe it thoroughly with 

HI 731318

 or a lint-

free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate

measurements, remove bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand for too long after reagent is added, as accuracy will

be affected.

• After the reading it is important to immediately discard the sample, otherwise the glass

might become permanently stained.