Preliminary examination:

Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each 

HI 718

 meter is supplied complete with:

• Two Sample Cuvettes and Caps
• Six packets of Iodine Reagent
• 1 x 1.5V AAA Battery
• Instruction Manual

HI 718

Technical specifications:

Range

0.0 to 12.5 ppm

Resolution

0.1 ppm

Accuracy

±0.1 ppm ±5% of reading @ 25 °C / 77 °F

Light Source

Light Emitting Diode @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Methods for the Examination
of Water and Wastewater, 18

th

 edition, DPD method.

The reaction between iodine and the reagent causes a
pink tint in the sample.

Environment

0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing

Battery Type

1 x 1.5V AAA

Auto-Shut off

After 10 minutes of non-use

Dimensions

81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)

Weight

64 g (2.25 oz.)

Functional description:

Errors and warnings:

Light High:

 There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.0”

 indicates that the sample absorbs

less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates

an over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be

replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.

1. Dust cover.

2. Cuvette with cap.

3. Cuvette holder.

4. Liquid Crystal Display.

5. Button.

For more details about spare parts and accessories
see “Accessories”.