Instruction Manual

www.hannainst.com

Dear Customer,
Thank you for choosing a Hanna Product.
Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information
for correct use of the kit. If you need additional information,
do not hesitate to e-mail us at [email protected].
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has

occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:

• HI 3834-0 Reagent, packets (100 pcs);
• 1 checker disc (containing a mirror, a mirror holder

and the 38039 disc);

• 2 glass vials with caps;
• 1 plastic pipette (3 mL).

Note:

Any damaged or defective item must be returned in
its original packing materials.

  

HI 38039

Iron Low Range

Test Kit with

Checker Disc

ISTR38039      09/00     PRINTED IN ITALY

Generally, ground and surface water contains no more than
1 mg/L (ppm) iron; but due to mining and industrial

drainage, higher levels of iron have been observed. Iron in
water appears to be more of a nuisance than a hazard.
The presence of iron can stain laundry and give water a
bittersweet taste.
The Hanna Test Kit determines the iron concentration in
water by conversion of the ferrous (Fe

2+

) state. The test is

fast and easy. The checker disc makes it simple to obtain
the iron level in water.

Note:

 mg/L is equivalent to ppm (parts per million).

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

SIGNIFICANCE AND USE

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

CHEMICAL REACTION

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

INSTRUCTIONS

Iron can exist as ferrous (Fe

2+

) or ferric (Fe

3+

) ions. The

Hanna Test Kit determines total iron levels in water via a
colorimetric method. First all ferric ions are reduced by
sodium sulfite to ferrous ions. Phenanthroline complexes with
ferrous ion to form an orange colored solution. The color
intensity of the solution determines the iron concentration.

READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Verify that the mirror, its holder and the disc are pre-

installed in the checker disc. If not, install them as
shown in the figure.

Range

0 to 1.0 mg/L (ppm) as Fe

Smallest Increment

0.02 mg/L (ppm) Fe

Analysis Method

Colorimetric

Sample Size

15 mL

Number of Tests

100

Case Dimensions

235x175x115 mm (9.2x6.9x4.5")

Shipping Weight

436 g (15.4 oz)

15 mL

The chemicals contained in this kit may be hazardous if

improperly handled. Read the relevant Health and Safety
Data Sheet before performing this test.

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

REFERENCES

REFERENCES

REFERENCES

REFERENCES

REFERENCES

1987 Annual Book of ASTM Standard

, Volume 11.01

Water (1), pages 531-535.

Standard Methods for the Examination of Water and
Wastewater

, 16

th

 Edition, pages 215-219.

• Using the plastic pipette, fill each glass

vial with 15 mL of sample (up to the
mark).

• Insert one of the vials into the left hand opening of the

checker disc. This is the blank.

• Add to the other

vial 1 packet of
HI 3834-0 re-
agent. Replace the
cap and shake to

mix.

• Wait for 3 minutes to allow reac-

tion to occur. This is the reacted
sample.

• Remove the cap and in-

sert the reacted sample
into the right hand open-
ing of the checker disc.

• Hold the checker disc so

that a light source illumi-
nates the samples from the
top.

• Keep the checker disc at a distance of 30-40 cm (12-

16") from the eyes to match the color while having a
uniform background (e.g. a white sheet). Rotate the
disc while looking at the color test windows and stop
when you find the color match. Read the value in the
result window directly in mg/L (or ppm) of  Iron.

For best results:

 Perform the reading three times and

take the average value (divide by 3 the sum of the
three numbers). Intensely colored samples will make
the color matching difficult and they should be

adequately treated before performing the test. Sus-
pended matter in large amounts should be removed
by prior filtration.

Caution:

 

Ultraviolet radiation may cause fading of colors.

When not in use, keep the disc protected from light,
in a cool and dry place.

Interferences:

 

Molybdate and Molybdenum above 50 ppm,

calcium above 10000 ppm (as CaCO

3

), magnesium

above 100000 ppm, chloride above 185000 ppm.