48
mV CALIBRATION
A two point calibration can be performed at 0 mV and 1800 mV.
• Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.
• Enter the calibration screen. Press
mV
function key.
• Set
0.0 mV
on the simulator.
• When the reading is stable and close to the selected calibration point, the
Confirm
function key is displayed.
• Press
Confirm
to confirm. The second calibration point of 1800 mV will be
displayed.
• Set
1800.0 mV
on the simulator.
• When the reading is stable and close to the selected calibration point, the
Confirm
function key is displayed.
• Press
Confirm
to confirm. The instrument returns to calibration screen.
• Press
ESC
to return to measurement mode.
Notes:
•If the reading is not close to the selected calibration point,
“WRONG”
tag will blink. Verify calibration condition or contact your vendor if
you cannot calibrate.
•Press CAL or
ESC
in any moment of the calibration process. The
instrument will return in the measurement mode.